1-800-663-4209 | Your feedback matters, take a quick survey Website Feedback

PR-25AP
-100 °C to 250 °C RTD Air Temperature Probes M12 Connector
Models In Stock
PR-25AP
From
$
195.50
Models In Stock

Product Overview
Accessories
M12 Cables with Field Mountable Connectors for Resistance Sensors (RTDs)
C$117.98
1 In Stock
M12 Cables with Field Mountable Connectors for Resistance Sensors (RTDs)
C$81.57
40 In Stock
M12 Cables with Field Mountable Connectors for Resistance Sensors (RTDs)
C$89.65
14 In Stock
M12 Cables with Field Mountable Connectors for Resistance Sensors (RTDs)
C$113.25
12 In Stock
Miniature Temperature Transmitter for Sanitary RTD Probes and Sensors
C$309.33
0 In Stock
Show Ratings & Reviews
Rated 5 out of
5
by
Chris Lanning from
Everything You Want from an RTD
Simple, accurate, and reliable. I used these for all clean air applications.
Date published: 2025-01-22
Rated 5 out of
5
by
Darren Youngblood from
My Go-To
Easy Order, Great Packaging. My Go-To source for Temp supplies
Date published: 2024-07-08
What is the 50% response time for these sensors?
Thank you for your inquiry! 50% Response is the time the resistance thermometer element needs in order to reach 50% of its steady state value. 90% Response is defined in a similar manner. These response times of elements are given for water flowing with 0. 2 m/s velocity and air flowing at 1 m/s. For more information on RTD response time, please see this Technical Reference from our UK site: https://www.omega.co.uk/temperature/z/resistanceelements.html
Date published: 2022-01-11
Is your "1/10 DIN" RTD actually "1/10*Class B" across their full operating range or only at 0°C? For example, is their accuracy ±0.13°C at 200°C (1/10 of Class B, which is ±1.30°C) or something else? If something else, please provide a formula or table.
Thank you for your inquiry. This is the tolerance equation for it: 1⁄10 DIN = ±1⁄10* (0.3 + 0.005*t)°C or 100.00 ±0.03 Ω at 0ºC
Date published: 2021-06-29
can you tell me the response time? the documentation just says fast. How fast?
Thank you for your inquiry! 50% Response is the time the resistance thermometer element needs in order to reach 50% of its steady state value. 90% Response is defined in a similar manner. These response times of elements are given for water flowing with 0. 2 m/s velocity and air flowing at 1 m/s. For more information on RTD response time, please see this Technical Reference from our UK site: https://www.omega.co.uk/temperature/z/resistanceelements.html
Application Engineer CK
Date published: 2022-01-10
Is this temperature probe submersible into liquids?
Thank you for your inquiry. This air probe is designed for measuring air only and cannot be submersed in liquids. We offer sealed probes that are designed for liquid submersion.
Date published: 2020-11-19
I know what a response time is and that they are determined either in water at 0. 2 m/s velocity and air flowing at 1 m/s. What is the response time for this sensor?
Thank you for your inquiry. Response Time would be less than 5 seconds.
Date published: 2022-01-25