Welcome to our new website, found an issue or bug? Please report it here

BB701-series
-18 °C to 149 °C Temperature Blackbody Calibrator
- Model BB701 Calibrates from -18 to 149°C (0 to 300°F )
- Portable Rugged Design
- Calibrates Infrared Pyrometers Quickly and Accurately
- Built in Digital PID Autotune Temperature Controller with Temperature Readout
BB701-series
From
$
5,714.35

Product Overview
- Accuracy ± 0.8 °C (± 1.44 °F)
- Accuracy Note ±0.8°C +1 digit (±1.4°F ) (worst case), ±0.3°C (±0.6°F) [over entire scale]
- Additional Features 1 Portable
- Calibration Reference Sensor Platinum RTD, 100 Ω, 0.00385
- Cavity Emissivity 0.95
- Control Mode PID Control
- Cool-Down Time Note 5 min, from ambient to -18°C (0°F)
- Data Interface RS232
- Display Type LED
- Frequency 50/60 Hz
- Height 19.5 mm
- Installation Category II
- Internal Sensing Element Platinum RTD, 100 Ω, 0.00385
- Length 36.5 mm
- Operating Temperature, Max 43 °C
- Operating Temperature, Min 4.4 °C
- Operating Temperature Note 4.4 to 29.4°C (40 to 85°F) [target plate reaches 0°F]; 4.4 to 43°C (40 to 110°F) [target plate reaches 11.8°F]
- Power Consumption 175 W
- Process Temperature, Max 149 °C
- Process Temperature, Min -18 °C
- RTD Accuracy ± 0.375 °C
- Stability ± 0.6°C
- Stability Note ± 0.6°C (± 1.2°F) or less
- Target Size 63.5
- Warm Up Time 5 min
- Warm-Up Time Note 5 min, from ambient to 149°C (300°F)
- Width 35.5 mm
The BB701 Blackbody Calibrator is a high performance, rugged, portable calibrator for infrared pyrometers. The BB701 Hot/Cold model has a range of -18 to 149°C (0 to 300°F). The BB701’s ability to provide a stable, repeatable cold calibration point allows the user to calibrate or test most infrared pyrometers quickly and accurately without having to prepare an ice bath. The target plate has a known emissivity, enabling the user to calibrate virtually any infrared pyrometer with a spot size diameter of 63.5 mm (2.5") or smaller. Both models come with an RS-232 computer interface which allows computer control of the setpoints for automatic test applications. An NIST traceable calibration certificate is also provided.
PLEASE NOTE: Items listed as "Stock" require 5 Business Days for NIST Certification and Order Processing
PLEASE NOTE: Items listed as "Stock" require 5 Business Days for NIST Certification and Order Processing
PDFs & Manuals
Show Ratings & Reviews
Rated 5 out of
5
by
Laboratory from
Calibration
Good product , good stability , great transaction recommended
Date published: 2021-02-13
Please provide details of the mating connector for the RTD. I know one came with the unit but it has long since been lost so we need to source a replacement
The connector needed to match the reference RTD probe connection is a TA3F.
Date published: 2024-12-23