1-800-663-4209 | Your feedback matters, take a quick survey Website Feedback

OSXL-SC-Series
-40 °C to 550 °C Temperature Bench Top Thermal Imager
Models In Stock
- For sales outside of the US and Canada
- Please order from our European Websites Directly.
- Designed Specifically for Thermal Benchtop Testing Applications
- Compact, Rugged, and Lightweight

Product Overview
- Accuracy ± 5 °C (± 9 °F)
- Base Mounting 4 × M3 thread mounting holes (bottom)
- Data Interface Ethernet
- Data Interface Note Ethernet: Control and image Type: Gigabit Ethernet Standard: IEEE 802.3 Connector Type: RJ45 Communication: GigE Vision version 1.2; Client API GenICam compliant Image Streaming: 14-bit signal linear/DDE, GigE Vision and GenICam compatible Power: Power over Ethernet, PoE IEEE 802.3af class 0 power Protocols: TCP, UDP, ICMP, IGMP, DHCP, GigEVision
- Detector Pitch 50 μm
- Digital Connector Type M12
- Digital Output 2 - 40 Vdc
- Digital Synchronization Connector Type M12
- Digital Synchronization Connector Type Note 12-pole M12 connector (shared with digital I/O and external power)
- Digital Synchronization Input Frame sync into control camera; 1×, non-isolated; LVC buffer @3.3V, “0” <0.8 V, “1” >2.0 V
- Digital Synchronization Output Frame sync out to control another Ax5 camera; 1×, non-isolated; LVC buffer @ 3.3V, ”0” = 24 MA maximum, “1” = -24 mA maximum
- Electrical Connection M12 4 pin Female
- Electrical Connection Note 12-pole M12 connector (shared with digital I/O and digital synchronization)
- Encapsulation IP 40 (IEC 60529)
- F-Number 1.25
- Field of view 44° x 36°
- Focal Length, Min 5 mm
- Focus Fixed
- Height 1.7 in
- Housing Material Magnesium,Aluminum
- Image Frequency 60 Hz
- Length 4.2 in
- Object Temperature, Max 550 °C
- Object Temperature, Min -40 °C
- Operating Temperature, Max 50 °C
- Operating Temperature, Min -15 °C
- Relative Humidity Note IEC 60068-2-30/24 h 95% relative humidity, 25 to 40°C (77 to 104°F) EMC: EN 61000-6-2 (immunity), EN 61000-6-3 (emission), FCC 47 CFR Part 15 Class B (emission)
- Shock 25 g
- Shock Note 25 g (IEC 60068-2-29)
- Spatial Resolution 10 mRad
- Spatial Resolution Note 10.0 mrad for 5 mm lens
- Spectral Response, Max 13 µm
- Spectral Response, Min 7.5 µm
- Spectral Response Note Uncooled VOX micro bolometer/7.5 to 13 μm
- Storage Temperature, Max 70 °C
- Storage Temperature, Min -40 °C
- Supply Power Note 12/24 Vdc, <2.5 W absolute maximum
- Thermal Sensitivity <0.05°C
- Thermal Sensitivity Note <0.05°C @ 30°C (86°F)/50 mK
- Time Constant 12 ms
- Tripod Mounting Optional with accessory T198349, base support
- Vibration 2 g
- Vibration Note 2 g (IEC 60068-2-6)
- Weight 0.2 kg
- Width 1.6 in
Thermal imaging cameras can be used for a wide variety of R&D applications. Until today using a thermal imaging camera was often the privilege for large R&D departments. With the extremely affordable OSXL-A35SC/ OSXL-A15SC and OSXL-A5SC thermal imaging systems are now bringing the advantages of thermal imaging to the test bench of every R&D engineer.
The OSXL-A35SC/OSXL-A15SC and OSXL-A5SC models are not only extremely affordable, they contain both the hardware and software for analyzing and verifying your R&D projects. You will be able to visualize and measure temperatures in a non-contact mode.
The OSXL-A35SC produces crisp thermal images of 320 x 256 pixels. Users that do not need this high image quality for their application can choose the OSXL-A15SC which produces thermal images of 160 x 128 pixels or the OSXL-A5SC which produces thermal images of 80 x 64 pixels.
The OSXL-A35SC, OSXL-A15SC, and OSXL-SC5 are affordable infrared camera kits designed specifically for thermal benchtop testing applications. The compact packaging makes the OSXL-A5SC a perfect fit for the benchtop and allows for deployment in locations where size constraints are critical. They are available in a variety of pixel resolutions and can meet the spatial resolution requirements of most applications.
Eliminate the guesswork, see heat patterns with the thermal imagery and extract temperature values from live or recorded imagery. This instrument has uncooled micro bolometer detector that is maintenance-free and provides excellent long wave imaging performance. The pixel resolution and optics are available in 80 x 64, 160 x 128, 320 x 256 pixel formats to achieve numerous fields of view. Versatile, compact, rugged, and lightweight with straightforward mounting that permits quick installation and easy movement for new application requirements. Plug-and-play compatibility, is an ideal system integration solution through GigE Vision and GenICam protocols, these cameras can be fully configured from a PC, allowingcamera control and image capture in real time. Fast data transfer and RJ45 GB Ethernet connection supply 14-bit images at frame rates as high as 60 Hz. Image and data acquisition can record thermal snapshots and movies with OSXL-ASC tools + recording and analysis software.
GigE Vision® is a new camera interface standard developed using the Gigabit Ethernet communication protocol. GigE Vision is the first standard video interface to allow for fast image transfer using low cost standard cables even over long distances. With GigE Vision, hardware and software from different vendors can interoperate seamlessly over GigE connections.
The goal of GenICamTM is to provide a generic programming interface for all types of cameras. Regardless of interface technology (GigE Vision, Camera Link®, 1394 DCAM, etc.) or features implemented, the Application Programming Interface (API) will always be the same. The GenICam protocol also makes it possible to use third party software with the camera. GenICam makes the OSXL-A35SC plug-and-play when used with software packages such as IMAQ Vision and Halcon.
The OSXL-A35SC/OSXL-A15SC and OSXL-A5SC models are not only extremely affordable, they contain both the hardware and software for analyzing and verifying your R&D projects. You will be able to visualize and measure temperatures in a non-contact mode.
The OSXL-A35SC produces crisp thermal images of 320 x 256 pixels. Users that do not need this high image quality for their application can choose the OSXL-A15SC which produces thermal images of 160 x 128 pixels or the OSXL-A5SC which produces thermal images of 80 x 64 pixels.
The OSXL-A35SC, OSXL-A15SC, and OSXL-SC5 are affordable infrared camera kits designed specifically for thermal benchtop testing applications. The compact packaging makes the OSXL-A5SC a perfect fit for the benchtop and allows for deployment in locations where size constraints are critical. They are available in a variety of pixel resolutions and can meet the spatial resolution requirements of most applications.
Eliminate the guesswork, see heat patterns with the thermal imagery and extract temperature values from live or recorded imagery. This instrument has uncooled micro bolometer detector that is maintenance-free and provides excellent long wave imaging performance. The pixel resolution and optics are available in 80 x 64, 160 x 128, 320 x 256 pixel formats to achieve numerous fields of view. Versatile, compact, rugged, and lightweight with straightforward mounting that permits quick installation and easy movement for new application requirements. Plug-and-play compatibility, is an ideal system integration solution through GigE Vision and GenICam protocols, these cameras can be fully configured from a PC, allowingcamera control and image capture in real time. Fast data transfer and RJ45 GB Ethernet connection supply 14-bit images at frame rates as high as 60 Hz. Image and data acquisition can record thermal snapshots and movies with OSXL-ASC tools + recording and analysis software.
GigE Vision® is a new camera interface standard developed using the Gigabit Ethernet communication protocol. GigE Vision is the first standard video interface to allow for fast image transfer using low cost standard cables even over long distances. With GigE Vision, hardware and software from different vendors can interoperate seamlessly over GigE connections.
The goal of GenICamTM is to provide a generic programming interface for all types of cameras. Regardless of interface technology (GigE Vision, Camera Link®, 1394 DCAM, etc.) or features implemented, the Application Programming Interface (API) will always be the same. The GenICam protocol also makes it possible to use third party software with the camera. GenICam makes the OSXL-A35SC plug-and-play when used with software packages such as IMAQ Vision and Halcon.
PDFs & Manuals
Show Ratings & Reviews
What are the export file formats?
Hi Foote,
Thank you for the inquiry. Please forward this question to TEMP@OMEGA.COM.
Link to full manual for this FLIR model;
https://assets.omega.com/manuals/M5230.pdf
Best regards
OMEGA
Date published: 2022-06-06
Are the measurements radiometric?
Thank you for the inquiry. Radiometric data is included when in 14 bit streaming mode.
Best regards
OMEGA
Date published: 2022-05-05