TXCO-50-Carbon-Monoxide-Transmitter

Carbon Monoxide Transmitter

  • 0 to 50 ppm Range
  • Compact and Inexpensive
  • Long Operating Life
  • NEMA 4 (IP65) Enclosure
TXCO-50-Carbon-Monoxide-Transmitter
From $ 1,035.20
Carbon Monoxide Transmitter TXCO-50-Carbon-Monoxide-Transmitter TXCO-50-Carbon-Monoxide-Transmitter
loading all models
Product Overview
  • Accuracy ± 2 %
  • Accuracy Detail ±2% FSD T = constant ±5% FSD 0>T>50°C (122ºF)
  • Additional Features 1 For Special Ranges Consult OMEGA
  • Calibration Using a known CO calibration gas that is NIST traceable
  • Electrical Output 4 to 20 mA
  • Enclosure Material Aluminum
  • Enclosure Ratings IP65, NEMA Type 4
  • Height 55 mm
  • Lead Length 0 ft
  • Length 80 mm
  • Load Resistance Typical 470 Ω at 18 Vdc maximum 700 Ω
  • Measuring Range 0 to 50 ppm CO
  • Operating Temperature, Max 50 °C
  • Operating Temperature, Min 0 °C
  • Pressure, Max 1 bar
  • Process Connection Size 1/4 in
  • Response Time 90% FSD at 20°C (68ºF) < 30 sec 90%
  • Supply Voltage 24 Vdc
  • Warm Up Time 2 min
  • Weight 1 kg
  • Width 75 mm

Description

The TXCO-50 Series of carbon monoxide transmitters is a simple, accurate and very economical way to measure CO in the 0 to 50 ppm levels. For special ranges please consult Omega.

The TXCO-50 sensor uses a special electro chemical cell to measure the CO concentration. The sensor meets the industrial requirements for accuracy, sensitivity and ease of use.

The transmitter comes fully calibrated and requires no calibration for two years. After two years the TXCO-50 can be calibrated with a CO gas. Typical sensor life is up to 5000 ppm hrs.

Specifications
Measuring Range: 0 to 50 ppm CO
Special Ranges: Consult OMEGA
Calibration: Using a known CO calibration gas that is NIST traceable
Accuracy:
     ±2% FSD T = constant
     ±5% FSD 0>T>50 °C (122 °F)
Response Time: 90% FSD at 20 °C (68 °F) < 30 sec 90%
Operating Temperature: 0 to 50 °C (32 to 122 °F)
Warm Up Time: 2 minutes
Pressure: 0.1 to 1 bar
Signal Output: 4 to 20 mA/DC loop power
Voltage: 24 Vdc
Load: Type 470 Ω at 18 Vdc maximum 700 Ω
Housing: NEMA 4 (IP65)
Dimensions: 80 L x 75 W x 55 mm D (3.15 x 2.95 x 2.16")
Enclosure: Painted aluminum
Connection: 1/4" tubing
Weight: 1.0 kg (2.2 lb)
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 24ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 52ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
en-US
en
Save Configuration