1-800-663-4209 | Your feedback matters, take a quick survey Website Feedback
CDCN-961
Conductivity, Resistivity or TDS Controller

Choose a Variant
Added to Your Shopping Cart
Configuration complete.
Please add to cart to keep it or exit the configuration
Please add to cart to keep it or exit the configuration
Failed to Save Configurations, Click on Configure Button Start Again
Product Specs
<< See all Product ModelsBrand |
![]() |
Accuracy | ± 1 % |
Cable Length | 5 m |
Calibration Note | Line Resistance Compensation: 0.0 to 20.0 Ω |
Cell Constant | K = 0.01, 0.1, 1, 10, Conductivity Electrode: K=1.0 |
Conductivity Measurement | 0.000 to 3.000, 30.00, 300.0 μS/cm |
Cutout Dimensions | 93.5 x 93.5 mm |
Data Interface | RS485 |
Diameter | 136 mm |
Display Type | LCD |
EMC Standard | EN 50081-1, EN 50082-1 |
Electrical Output | 4 to 20 mA |
Enclosure Ratings | IP65, NEMA Type 4X |
Frequency | 1 to 999 hours 1 to 999 seconds(Frequency of Cleaning Time) |
Height | 108 mm |
Impedance | 600 Ω |
Materials | Graphite |
Measurement Range Note | Conductivity: 0.000 to 3.000, 30.00, 300.0 μS/cm Resistivity: 0.3 to 30.0 MΩcm TDS: 0.000 to 3.000, 30.00, 300.0 ppm, Test Range: 0 to 50.0 mS/cm |
Number of Relays | 3 |
Operating Humidity | 10 to 95% RH |
Operating Temperature, Max | 50 °C |
Operating Temperature, Min | -10 °C |
Output Detail | 4 to 20 mA/600 Ω |
Process Temperature, Max | -0.5 °C |
Process Temperature, Min | 130 °C |
Relay Note | 3 SPST/250V 1A, Control Hysteresis Band: 0.1 to 10% FS |
Relay Output | SPST |
Resistivity Measurement Range | 0.3 to 30.0 MΩcm |
Resolution | 0.001/0.01/0.1 μS/cm |
Supply Voltage | 220 Vac |
Temperature Accuracy | ± 0.5 °C |
Temperature Resolution | 0.1 °C |
Temperature Sensor Type | 1000 Ω RTD |
Weight | 1650 g |
Width | 100 mm |
See All Specs |