Skip to search Skip to main content Skip to navigation

Welcome to our new website, found an issue or bug? Please report it here

We noticed you are from . Visit your local site for regional offers and live support.

CL532

Milliamp Loop Calibrator

$2,503.81

Item# CL532

Volume discounts available
Volume Discount Schedule
Quantity Price each
1-9 $2,503.81
10-24 $2,378.62
25-49 $2,328.54
50-99 $2,228.39
100-199 $2,178.31
200+ $2,128.24

In Stock

Lead Time (If not in stock): 5 weeks

Product Specs

<< See all Product Models
Brand
2-Wire Transmitter Simulate Adjust loop current from 0 to 24 mA or -25 to 125% when powered by an external supply; LOW SUPPLY flashes to indicate a misconnection.
Accuracy ± 0.012 %
Accuracy Note ±0.012% rdg + 0.004 mA
Battery Size and Type Alkaline 9V
Battery life 18hr
Display Type High-Contrast Display with Bar Graph
Electrical Connection Alligator Clips
Height 180 mm
Lead Length 610 mm
Length 56 mm
Loop Supply Limits 2 to 60 Vdc; Drive Capability: 1200 Ω at 20 mA for life of battery; Burden: <2V at 20 mA
Measuring Range Read Current: Read 0 to 24 mA or -25 to 125%; CURRENT LIMITED flashes at >24 mA, protecting devices in the loop.; Read DC Voltage: Read 0 to 60 Vdc with 2x overrange ability and flashing OVERRANGE indicator; Power Measure: Power the loop and read 0 to 24 mA or -25 to 125%; CURRENT LIMITED flashes at >24 mA, protecting devices in the loop.
Miscellaneous EZ-Check Switch: Gives instant access to 4 and 20 mA or other user-set span limits; in read modes, he EZ-Check switch provides a max/min log of reading; EZ-Step Button: Allows one-handed stepping up and down the span in 6 different step sizes; Adjusts output by 0.001 mA (0.01%); EZ-Dial Knob: push and turn for 100x faster adjustment—by 0.100 mA (1.00%)
Operating Temperature, Max 60 °C
Operating Temperature, Min -20 °C
Over-Voltage Protection 120 Vrms for 30 seconds, 240 Vrms for 15 seconds
Overcurrent Protection 24 mA DC
Relative Humidity 10% ≤RH ≤90% [0 to 35°C (32 to 95°F)], non-condensing; 10% ≤RH ≤70% [35 to 60°C (95 to 140°F)], non-condensing
Source Current Source 0 to 24 mA or -25 to 125% into ≤1200 Ω; a flashing HIGH Ω indicator appears if improperly connected
Storage Temperature, Max 60 °C
Storage Temperature, Min -30 °C
Temperature Effect ±50 ppm/°C
Weight 326 g
Width 84 mm
See All Specs Show Less