Welcome to our new website, found an issue or bug? Please report it here

RH820-Series
Handheld Humidity, Temperature, and Dew Point Meters
Models In Stock
- USB Port Provides Excellent User Interface (RH820U)
- Wireless 2-Way Transmission (RH820W)
- 2 Temperature Sensors (Internal NTC and External K-Type) and Humidity Sensor Together
- Wet/Dry Bulb, Dew Point and %RH
RH820-Series
From
$
309.95
Models In Stock

Product Overview
- Battery Indication A low battery indication is displayed when the battery voltage drops below the operating level
- Display Type 3.5 Digit LCD
- Height 160 mm
- Humidity Sensor Capacitive humidity
- Hysteresis ± 1%
- Measuring Temperature Range, Max 1000 °C
- Measuring Temperature Range, Min -50 °C
- Operating Temperature, Max 50 °C
- Operating Temperature, Min 0 °C
- Process Temperature, Max 60 °C
- Process Temperature, Min 0 °C
- Relative Humidity Accuracy ± 2.5% RH
- Relative Humidity Accuracy Note ± 2.5% at 25°C (77°F), 10 to 90% RH; ± 5% at 25°C (77°F), 0 to 10% RH, 90 to 100% RH
- Response Time 60 sec
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -20 °C
- Storage Temperature Note -20 to 60°C (-4 to 140°F), 0 to 70% RH with battery removed from meter
- Supply Voltage 6 Vdc
- Temperature Accuracy ± 0.1% rdg +1 °C
- Temperature Accuracy Note ± (0.1% rdg +1°C) on -50 to 1000°C; ± (0.1% rdg +2°F) on -58 to 1832°F
- Temperature Resolution 0.1 °C
- Temperature Sensor Accuracy ± 2 °C
- Temperature Sensor Accuracy Note ± 2°C on 0 to 10°C, ± 0.5°C on 10 to 45°C; ± 2°C on 45 to 60°C; ± 4°F on 32 to 50°F, ± 1°F on 50 to 113°F; ± 4°F on 113 to 140°F
- Temperature Sensor Type Thermistor
- Width 83 mm
High performance humidity handheld meters available with USB or wireless connectivity. They offer a double display to show the temperature and humidity from built in sensors and can alternately display data from a separate Type K thermocouple connected via SMP type miniature connector. Save function allows storage of up to 256 readings (no time-stamp) to internal non-volatile memory. Software for USB and wireless versions is included to allow internal, external temperature, and RH data to be recorded by a PC. Interface cable is included with the USB version; receiver with the wireless version.
SPECIFICATIONS
Display: 3.5" digit LCD with a maximum reading of 1999
Low Battery Indication: The "Low Battery Icon" is displayed when the battery voltage drops below the operating level
Operating Environment: 0 to 50°C (32 to 122°F) at <80% RH
Storage Environment: -20 to 60°C (-4 to 140°F), 0 to 70% RH with battery removed from meter
Accuracy: Stated accuracy at 23°C ± 5°C, <75% RH
Battery: Four 1.5V "AAA" (included)
Dimensions: 160 H x 83 W x 38 mm D (6.3 x 3.3 x 1.5")
Weight: Approx 265 g (0.6 lb) including batteries
Temperature Range: -50 to 1000°C (-58 to 1832°F)
Accuracy: ± (0.1% rdg +1°C) on -50 to 1000°C; ± (0.1% rdg +2°F) on -58 to 1832°F
Temperature Sensor: Thermistor temperature sensor
Range: 0 to 60°C (32 to 140°F)
Resolution: 0.1°C/°F
Accuracy: ± 2°C on 0 to 10°C, ± 0.5°C on 10 to 45°C; ± 2°C on 45 to 60°C; ± 4°F on 32 to 50°F, ± 1°F on 50 to 113°F; ± 4°F on 113 to 140°F
Relative Humidity Sensor: Capacitive humidity sensor
Range: 0 to 100% RH, noncondensing
Accuracy: ± 2.5% at 25°C (77°F), 10 to 90% RH; ± 5% at 25°C (77°F), 0 to 10% RH, 90 to 100% RH
Sensor Response Time for 90% of Total Range: 60 sec typical
Sensor Hysteresis (Excursion of 10% to 90% to 10% RH): ± 1% RH typical
RH820W Frequency Range: 910~920 mHz, low current consumption less than 1 mA, the transmitting distance can reach 25M
SPECIFICATIONS
Display: 3.5" digit LCD with a maximum reading of 1999
Low Battery Indication: The "Low Battery Icon" is displayed when the battery voltage drops below the operating level
Operating Environment: 0 to 50°C (32 to 122°F) at <80% RH
Storage Environment: -20 to 60°C (-4 to 140°F), 0 to 70% RH with battery removed from meter
Accuracy: Stated accuracy at 23°C ± 5°C, <75% RH
Battery: Four 1.5V "AAA" (included)
Dimensions: 160 H x 83 W x 38 mm D (6.3 x 3.3 x 1.5")
Weight: Approx 265 g (0.6 lb) including batteries
Temperature Range: -50 to 1000°C (-58 to 1832°F)
Accuracy: ± (0.1% rdg +1°C) on -50 to 1000°C; ± (0.1% rdg +2°F) on -58 to 1832°F
Temperature Sensor: Thermistor temperature sensor
Range: 0 to 60°C (32 to 140°F)
Resolution: 0.1°C/°F
Accuracy: ± 2°C on 0 to 10°C, ± 0.5°C on 10 to 45°C; ± 2°C on 45 to 60°C; ± 4°F on 32 to 50°F, ± 1°F on 50 to 113°F; ± 4°F on 113 to 140°F
Relative Humidity Sensor: Capacitive humidity sensor
Range: 0 to 100% RH, noncondensing
Accuracy: ± 2.5% at 25°C (77°F), 10 to 90% RH; ± 5% at 25°C (77°F), 0 to 10% RH, 90 to 100% RH
Sensor Response Time for 90% of Total Range: 60 sec typical
Sensor Hysteresis (Excursion of 10% to 90% to 10% RH): ± 1% RH typical
RH820W Frequency Range: 910~920 mHz, low current consumption less than 1 mA, the transmitting distance can reach 25M
PDFs & Manuals
Show Ratings & Reviews
Rated 1 out of
5
by
Anonymous from
Purchasing Agent
We got sent ones that the cert for calibration was expired already.
Date published: 2020-10-28
Rated 4 out of
5
by
MME Forensic from
Mostly good, one disapppointment
I like all the features and know that I'll likely get a lot of use out of it. The disappointment is the first time I used it and saved 240 data points 80% of the file was corrupted when I downloaded to Excel. Only the first 48 data points were good. The rest were trash. Maybe I did something wrong, so I'll keep trying.
Date published: 2019-03-16
can I measure relative humidity inside an oven where the temperature could be up to 95C?
Hi Gani,
Thank you for the inquiry. The sensor is located at the top of the handheld and the max operating temperature range is limited to 50 C.
Best regards
OMEGA
Date published: 2024-03-18
Is this item available with a calibration certificate?
Thank you for your inquiry. This item is not available with calibration.
Date published: 2020-07-08