Welcome to our new website, found an issue or bug? Please report it here

OSAT-Series
Intrinsically Safe IR Sensor for Hazardous Locations
Models In Stock
OSAT-Series
From
$
842.12
Models In Stock

Product Overview
- Accuracy Note ± 1°C or 1%, whichever isgreater
- Cable Length Note 5, 10 or 25 m (16.4, 32.8, 82') as standard (custom lengths also available) Weight: 475 g (1 lb), wi
- Current Draw 25 mA
- Emissivity, Max 1
- Emissivity, Min 0.2
- Emissivity Note 0.20 to 1.00 (pre-set to 0.95)
- Emissivity Setting Method USB
- Enclosure Ratings IP65, NEMA Type 4
- Housing Diameter 20 mm
- Housing Length 150 mm
- Housing Material 316 Stainless Steel
- Operating Temperature, Max 70 °C
- Operating Temperature, Min 0 °C
- Relative Humidity, Max 0.95
- Repeatability ± 0.5%
- Repeatability Note ± 0.5°C or 0.5%, whichever is greater
- Response Time 240 ms
- Response Time Note 240 ms (90% response)
- Sensing head Mounting 20 mm
- Spectral Response, Max 14 μm
- Spectral Response, Min 8 μm
- Supply Voltage 12 to 24 Vdc
- Weight 475 g
- Weight Note 475 g (1 lb), with 5 m (16.4') cable
Show Ratings & Reviews
Is this rated for an environment containing alkali metals (and their vapors)?
Thank you for the inquiry. Please forward this request to TEMP@OMEGA.COM.
Best regards
OMEGA
Date published: 2023-09-09
Why is the operating range 0 deg C to 70 deg C? What happens below 0 deg C ambient? I understand the upper operating limit 70 deg C.
The operating range defines the range were the IR sensor will meet the specifications. Outside this range the specifications will not be guaranteed and long term performance could be compromised.
Date published: 2025-01-13
Does the osat have to be used with the MTL 5541 to be intrinsically safe or can it be loop powered from a DCS?
Hi mjohn,
Thank you for the request. The OSAT requires use of a certified intrinsically safe isolator to be considered intrinsically safe. The OSAT cannot be powered direct from a common DCS and still be considered intrinsically safe.
Best regards,
OMEGA
Date published: 2022-12-08
RoHS 2015/863 - View Certificate
CE Approved
UKCA
ATEX