- A/D Conversion 24-bit Sigma Delta
- Cutout Dimensions 92 x 92 mm
- Data Interface RS232, RS485
- Data Interface Note RS232 and RS485 (User Selectable) Up to 115.2k baud Modbus Protocol (RTU)
- Diameter 135 mm
- Display Size 21 mm (.82 in)
- Display Type 4 Digit LED
- Display Type Note 4-digit, 7-segment LED; red, 21 mm (0.83), 1 to 40s
- Form factor 1/4 DIN
- Height 95 mm
- Input Control Signal 4 to 20 mA, 0 to 1 V
- Input Type Thermocouple, RTD, mA, Voltage
- Input Type Note Input Types: TC: J, K, T, E, R, S, B, C, N RTD: 2 or 3 wire PT100 0.00385 CU10 0.00427 NI120 0.00672 1RTD 3 Wire 6 channels for all models
- Material Construction Option N/A
- Measuring Range Process Voltage: 0 to 1000 mV Process Current: 0 to 24.00 mA Type J: -150 to 1200°C Type K: -150 to 1372°C Type T: -150 to 400°C Type E: -150 to 1000°C Type R: -50 to 1788°C Type S: -50 to 1768°C Type B: 150 to 1820°C Type C: 0 to 2320°C Type N: -150 to 1300°C Pt RTD: 200 to 850°C Cu RTD: -200 to 260°C Ni RTD: -80 to 260°C
- Number of Relays 2
- Operating Temperature, Max 70 °C
- Operating Temperature, Min -20 °C
- Panel Depth 140 mm
- Power Consumption 3 W
- Resolution 1°, 1 mV, 0.01 mA
- Retransmission/Analog Output 2 Relays
- Sampling Rate 2.5 Hz (2.5 updates/second)
- Scan Rate 400 ms
- Storage Temperature, Max 85 °C
- Storage Temperature, Min -40 °C
- Supply Power Note Input Power: AC Models: 120/240 Vac; 50/60 Hz DC Models: 9 to 36 Vdc 3 watts maximum
- Thermocouple Type J, K, T, E, R, S, B, C, N
- Weight 725 g
- Width 95 mm
The DP606A and DP612A Series is our newest line of affordable multi-channel panel meters. These models offer increased accuracy, wider operating temperatures and increased flexibility over the CN606 and CN612 models that they replace. The input, power and communications are also now isolated allowing for easier deployment and more reliable operation. The rugged extruded aluminum ¼ DIN case combines with the improved isolation to provide an extremely durable product.
These meters can read up to 12 sensors and include user selectable RS232 or RS485 communications using Modbus®. Two user configurable alarm relays are standard on all models. The universal input allows user selection between Thermocouple (9 Types), RTD 2-Wire, RTD 3-Wire, 4 to 20 mA and 1000 mV input options. These input selections are firmware selectable through the front panel buttons or Modbus serial commands. No dip switch setting changes are required.
Excellent fit for OEM instrumentation, automotive and aerospace test stand, factory automation, and power generation applications.
Note: While the inputs are universal and can be changed via firmware (no dip switches), all 6 or 12 inputs on a given unit must be set to the same type of input for operation.
Software: This multi-zone panel meter is most easily configured with Omega Sync, a software tool that enables users to configure device runtime parameters, view process values, and seamlessly export data. At the moment, Omega Sync supports the CN616A and DP606A/DP612A devices and we look to expand its reach to future products. For long-term data storage, our experts recommend using Omega Enterprise Gateway (OEG) that provides a powerful data historian, alarming, visualization and OPC-UA integration.
Specifications
Number of Zones: 61 or 12
1RTD 3 Wire 6 channels for all models
Display: 15 mm 7-seg LED
Display Time: 1 to 40s
Scan Time: 400 mS
Input Types:
TC: J, K, T, E, R, S, B, C, N
RTD: 2 or 3 wire
PT100 0.00385
CU10 0.00427
NI120 0.00672
Process: 4 to 20 mA, 0 to 1V
Resolution 1º, 1 mV, 0.01 mA
Accuracy: See Chart Below
A2D: 24-bit Sigma Delta
Communications:
RS232 and RS485 (User Selectable)
Up to 115.2k baud
Modbus Protocol (RTU)
Input Power:
AC Models: 120/240 Vac; 50/60 Hz
DC Models: 9 to 36 Vdc
3 watts maximum
Isolation per 1 Min
Power to Input:
AC Units: 2.3k Vac (3.2k Vdc)
DC Units: 1.5k Vdc
Power to Alarm Relay: 2.3k Vac (3.2k Vdc)
Power to Serial: 2.3k Vac (3.2k Vdc)
Input to Serial: 500 Vac (720 Vdc)
Alarm Relays:
AC Models:
240 Vac max
5A max load
DC Models:
36 Vdc max
3A max load
De-energizes on Alarm Latching or Non-Latching High, Low or Hi/Low
Operating Temp: -20 to 70ºC (-4 to 158ºF)
Storage Temp: -40 to 85ºC (-40 to 185ºF)
Panel Cutout: ¼ DIN 92 x 92 mm (3.6 x 3.6")
Dimensions: 95 H x 95 W x 135 Dia mm (3.7 x 3.7 x 5.3")
Min Rear Panel Depth: 140 mm (5.5")
Weight: 725 g (26 oz)
Approvals
ECCN: EAR99
Safety: 2014/35/EU
EMC: 2014/30/EU
RoHS: 2011/65/EU
UL File No.: E209855
UL / cUL Listed
These meters can read up to 12 sensors and include user selectable RS232 or RS485 communications using Modbus®. Two user configurable alarm relays are standard on all models. The universal input allows user selection between Thermocouple (9 Types), RTD 2-Wire, RTD 3-Wire, 4 to 20 mA and 1000 mV input options. These input selections are firmware selectable through the front panel buttons or Modbus serial commands. No dip switch setting changes are required.
Excellent fit for OEM instrumentation, automotive and aerospace test stand, factory automation, and power generation applications.
Note: While the inputs are universal and can be changed via firmware (no dip switches), all 6 or 12 inputs on a given unit must be set to the same type of input for operation.
Software: This multi-zone panel meter is most easily configured with Omega Sync, a software tool that enables users to configure device runtime parameters, view process values, and seamlessly export data. At the moment, Omega Sync supports the CN616A and DP606A/DP612A devices and we look to expand its reach to future products. For long-term data storage, our experts recommend using Omega Enterprise Gateway (OEG) that provides a powerful data historian, alarming, visualization and OPC-UA integration.
Specifications
Number of Zones: 61 or 12
1RTD 3 Wire 6 channels for all models
Display: 15 mm 7-seg LED
Display Time: 1 to 40s
Scan Time: 400 mS
Input Types:
TC: J, K, T, E, R, S, B, C, N
RTD: 2 or 3 wire
PT100 0.00385
CU10 0.00427
NI120 0.00672
Process: 4 to 20 mA, 0 to 1V
Resolution 1º, 1 mV, 0.01 mA
Accuracy: See Chart Below
A2D: 24-bit Sigma Delta
Communications:
RS232 and RS485 (User Selectable)
Up to 115.2k baud
Modbus Protocol (RTU)
Input Power:
AC Models: 120/240 Vac; 50/60 Hz
DC Models: 9 to 36 Vdc
3 watts maximum
Isolation per 1 Min
Power to Input:
AC Units: 2.3k Vac (3.2k Vdc)
DC Units: 1.5k Vdc
Power to Alarm Relay: 2.3k Vac (3.2k Vdc)
Power to Serial: 2.3k Vac (3.2k Vdc)
Input to Serial: 500 Vac (720 Vdc)
Alarm Relays:
AC Models:
240 Vac max
5A max load
DC Models:
36 Vdc max
3A max load
De-energizes on Alarm Latching or Non-Latching High, Low or Hi/Low
Operating Temp: -20 to 70ºC (-4 to 158ºF)
Storage Temp: -40 to 85ºC (-40 to 185ºF)
Panel Cutout: ¼ DIN 92 x 92 mm (3.6 x 3.6")
Dimensions: 95 H x 95 W x 135 Dia mm (3.7 x 3.7 x 5.3")
Min Rear Panel Depth: 140 mm (5.5")
Weight: 725 g (26 oz)
Approvals
ECCN: EAR99
Safety: 2014/35/EU
EMC: 2014/30/EU
RoHS: 2011/65/EU
UL File No.: E209855
UL / cUL Listed
Measurement Types and Temperature Ranges | Operating Temperature Range | ||||
---|---|---|---|---|---|
Input Type | Description | Range | "Accuracy (25ºC)" | "Accuracy (0 to 50ºC)" | "Accuracy (-20 to 70ºC)" |
Process | Process Voltage | 0 to 1000 mV | ±1 mV | ±1 mV | ±1 mV |
Process Current | 0 to 24.00 mA | ±10 µA | ±10 µA | ±10 µA | |
J Type T/C | J Iron-Constantan | -150 to 0°C | ±1.0°C | ±2.0°C | ±6.0°C |
0 to 1200°C | ±1.0°C | ±1.0°C | ±2.0°C | ||
K Type T/C | CHROMEGATM-ALOMEGATM | -150 to 0°C | ±1.0°C | ±2.0°C | ±5.0°C |
0 to -1372°C | ±1.0°C | ±1.0°C | ±2.0°C | ||
T Type T/C | Copper-Constantan | -150 to 0°C | ±1.0°C | ±2.0°C | ±7.0°C |
0 to 400°C | ±1.0°C | ±1.0°C | ±2.0°C | ||
E Type T/C | CHROMEGATM-Constantan | -150 to 0°C | ±1.0°C | ±2.0°C | ±5.0°C |
0 to 1000°C | ±1.0°C | ±1.0°C | ±2.0°C | ||
R Type T/C | Pt/13%Rh-Pt | -50 to 0°C | ±1.0°C | ±2.0°C | ±6.0°C |
0 to 1788°C | ±1.0°C | ±1.0°C | ±2.0°C | ||
S Type T/C | Pt/10%Rh-Pt | -50 to 0°C | ±1.0°C | ±2.0°C | ±5.0°C |
0 to1768°C | ±1.0°C | ±1.0°C | ±2.0°C | ||
B Type T/C | 30%Rh-Pt/6%Rh-Pt | 150 to 700°C | ±1.0°C | ±2.0°C | ±3.0°C |
700 to 1820°C | ±1.0°C | ±1.0°C | ±1.0°C | ||
C Type T/C | 5%Re-W/26%Re-W | 0 to 2320°C | ±1.0°C | ±1.0°C | ±3.0°C |
N Type T/C | Nicrosil-Nisil | -150 to 0ºC | ±1.0°C | ±2.0°C | ±5.0°C |
0 to 1300°C | ±1.0°C | ±1.0°C | ±2.0°C | ||
RTD 2/3 Wire | Pt, 0.00385, 100? | -200 to 850°C | ±1.0°C | ±1.0°C | ±1.0°C |
RTD 2/3 Wire | Cu, 0.00427, 10? | -200 to 260°C | ±1.0°C | ±1.0°C | ±1.0°C |
RTD 2/3 Wire | Ni, 0.00672, 120? | -80 to 260°C | ±1.0°C | ±1.0°C | ±1.0°C |
Show Ratings & Reviews
Rated 5 out of
5
by
jastew358 from
Universal 6/12 panel meter
This product is a great addition to our machine safety; easy to use and consistent
Date published: 2020-12-29
I was hoping to connect the meter to a pc to log data in conjunction with the OEM software. Is this possible? If so, where could I find the wiring diagram for the serial connection? I've looked in the manual but cannot find any specifics for this wiring.
The 3 terminal connector marked serial is where you make your communication connection. Pin #1 code B is your Transmit (TX), pin #2 code A is your receive (RX) and pin #3 is the RTN connection. You can connect directly to your PC's Comm port or use a OM-CONV-USB and go into a USB input.
Date published: 2023-12-12
Are there any examples of how to communicate with modbus using Labview? (i.e. an instruction to read temperature through visa)
Thank you for the inquiry. We do not have labview examples unfortunately. But the modbus register assignments and details are given in the manual;
https://assets.omega.com/manuals/DP600A_Manual.zip
Best regards
OMEGA
Date published: 2020-01-14
Should the controller automatically read multiple temperatures from multiple thermocouples after installation or is there a setting I am missing?
On the DP606A there should be one temperature displayed per active zone with a Thermocouple input. Function 2 sets which zones are active. If you have further questions contact the Temperature Engineering group at temp@omega.com.
Date published: 2022-09-22
I am trying to communicate between my PC and this controller through labview, but have trouble sending data. Are there any examples of modbus instructions using visa available? (i.e. how to read temperature or change between Fahrenheit/Celsius)
Thanks again 1anh21.
We do not have labview examples but modbus details are provided in main manual linked here;
https://assets.omega.com/manuals/DP600A_Manual.zip
Best regards
OMEGA
Date published: 2020-01-14
looking for DP612A Panel meter Operating and set up instruction <we have purchased last week ,
Thank you for your inquiry! You can find the manual in the "Manuals" tab under Product Specs.
Date published: 2021-07-08
Would like to know if there is a Modus register that can be used to stop the automatic temperature values loop and just show the temperature value of a specific sensor.
Hi guillermoe
Thank you for the inquiry. The necessary commands were not brought out to the modbus registers.
Best regards
OMEGA
Date published: 2021-09-28
Why even use RTD, the accuracy still ±1 degree C, RTD itself should be within ±0.5 degree C between 0 to 100 C, can the Controller maintain the temperature measurement within 0.5 degree C accuracy?
Thank you for your inquiry. The accuracy for the RTD Input is ±1°C. You still need to add the accuracy of the RTD itself to the overall accuracy.
Date published: 2022-05-16