Welcome to our new website, found an issue or bug? Please report it here

LCEB
Low Profile, Mini Beam Load Cell for Low Capacities
Models In Stock
- Very High Precision
- Low Profile - 1" Height
- High Repeatability
- 100% Creep Tested
LCEB
From
$
299.08
Models In Stock

Product Overview
- Cable Length 5 ft
- Calibration Option NIST
- Creep in 20 min ± 0.15%
- Electrical Connection Cable
- Hysteresis ± 0.02%
- Input Resistance 350 +50/-3.5 Ω
- Load Cell Type Bending Beam
- Operating Temperature, Max 90 °C
- Operating Temperature, Min -55 °C
- Output Resistance 350 ± 3.5 Ω
- Overload, Max ± 400% FS
- Process Temperature, Max 65 °C
- Process Temperature, Min -15 °C
- Repeatability ± 0.01%
- Safe Overload 150% FS
- Supply Power Note 10 Vdc, 15 Vdc maximum
- Supply Voltage 10 to 15 Vdc
- Thermal Span Effect ± 0.0008% FS/°F
- Thermal Zero Effect ± 0.0015% rdg/°F
- Wiring/Connector Table Green: (+) Output; White: (-) Output; Black: (-) Input; Red: (+) Input
- Zero/Balance Error ± 1%
Model LCEB load cells offer exceptional accuracy in compression in controlled environments (in general anywhere a digital readout is used). Their high accuracy, small size, low creep, and excellent temperature compensation make the LCEB series an excellent choice for applications such as force testing, thrust measurements, check weighing, and scale applications.
Show Ratings & Reviews
Hi, Could we use 5vdc excitation with this product? Does it affect the precision or the linearity spec ? Thanks
You can use 5VDC without effecting the accuracy or linearity. The only difference would be that the output signal of the load cell would be 0 to 15mV instead of 0 to 30mV
Date published: 2023-06-09
Would this LCEB series load cell hold up to tests with very high relative humidity? (10-100%) Would it's measuring capabilities be compromised?
The LCEB has an IP65 environmental rating. It should be able to handle dust and water ingress
Date published: 2024-02-19