Welcome to our new website, found an issue or bug? Please report it here

PHE4200-Series
Clear Epoxy-bodied, Gel-filled Combination Alpha® Electrodes
- Five Electrode Styles
- Covers Entire 0 to 14 pH Range
- Clear Epoxy Body Construction
- Fast Responding: 95% in Less than 1 Second
PHE4200-Series
From
$
125.30

Product Overview
- Enclosure Material Epoxy, PP
- Impedance 50 MΩ
- Measurement Type pH
- PH Measurement Range 0 to 14 pH
- Process Connection Size N/A
- Process Connection Type N/A
- Process Temperature, Max 140 °C
- Process Temperature, Min 0 °C
- Response Time 95% in Less than 1 Sec
- Sensor Type N/A
- Temperature Sensor Type N/A
OMEGA Engineering is pleased to offer a complete line of laboratory pH electrodes. The PHE-4200 Series are gel-filled, combination pH electrodes with a unique clear epoxy body design. With five electrode styles to choose from, you’re sure to find one suitable for your application. OMEGA ® PHE-4200 electrodes offer reliable quality at a reasonable cost. The electrodes measure the entire pH range from 0 to 14 pH units at 0 to 100°C (32 to 212°F) and feature an Ag/AgCI reference and polypropylene liquid junction. Each electrode comes in a soaker storage bottle to keep the electrode moist and ready for use. Shelf life is guaranteed for one year, and the OMEGA quality certificate provided assures that these electrodes meet the most stringent quality control requirements.
PDFs & Manuals
Show Ratings & Reviews
Rated 5 out of
5
by
Procurement Officer from
10/10 always repurchase
Durable and fit for purpose. Our company uses it to test glues and it is easy to clean without damaging any part of the component.
Date published: 2023-04-20
Are there any installation arrangement requirements to be aware of? Does entire length need to be submerged? Does orientation matter?
Thank you for your question?
pH probes should be mounted vertically or at least 15deg above the horizontal plane in the direction of flow. At a minimum the bulb at the end of the probe needs to be fully submerged for an accurate reading.
Date published: 2019-07-05
How long is the cable on this electrode?
Thank you for your inquiry about the pH probes. It has 2.5 feet of cable with a BNC connector.
Date published: 2022-12-01
Will this probe work in D ionized water?
Thank you for your inquiry. We would recommend the PHE-4202-LC for DI water. It is the same as the PHE-4202 but for low conductivity fluids. This will be a custom probe so please contact us at flow@omega.com for price and lead time.
Date published: 2019-11-17