Welcome to our new website, found an issue or bug? Please report it here

OS36-10-K
Built-in Air Purge IR Sensor with Thermocouple Output
- Built-in Air Purge
- K Thermocouple Output Signal
- Use With Standard Thermocouple Indicators and Controllers
- Self-Powered

Product Overview
The OS36-10 Series provides precise temperature control for optimum product quality and maximum throughput. Like conventional thermocouples, all IR thermocouples do not require external power and plug into your controller, PLC, datalogger, recorder or data acquisition system. They simply measure temperature without contact. The internal air purge passages provide a highly efficient method of maintaining optical cleanliness in harsh environments.
SPECIFICATIONS
Range: -45 to 1370°C (-50 to 2500°F)**
Field-of-View: 6° (10:1)
Minimum Spot Size: 20 mm (0.8")
Output Impedance: 4 to 8 kΩ (varies by model)
Operating Ambient Temperature Range: -18 to 85°C (0 to 185°F)
Air Purge: Built-in
Ambient Temperature Compensation: With air flow of 1.1 SCFM (20 psig) maximum ambient is 204°C (400°F)
Dimensions: 9.55 L x 3.49 cm Dia. (3.76 x 1.375"); 0.9 m (36") PFA coated lead wires
Weight: 184 g (6.5 oz) with cable
**See range chart for linear ranges. Measurement is practical beyond thermocouple table values by use of radiation laws. Polynomials available upon request.
Range Code | Temperature Range (2% Accuracy) |
---|---|
50F | -18 to 27°C (0 to 80°F) |
80F | 10 to 49°C (50 to 120°F) |
98.6F | 16 to 39°C (60 to 102°F) |
140F | 25 to 80°C (80 to 180°F) |
180F | 70 to 104°C (160 to 220°F) |
240F | 95 to 130°C (200 to 270°F) |
280F | 125 to 155°C (260 to 310°F) |
340F | 145 to 175°C (290 to 350°F) |
440F | 170 to 250°C (340 to 480°F) |
Range: -45 to 1370°C (-50 to 2500°F)**
Field-of-View: 6° (10:1)
Minimum Spot Size: 20 mm (0.8")
Output Impedance: 4 to 8 kΩ (varies by model)
Operating Ambient Temperature Range: -18 to 85°C (0 to 185°F)
Air Purge: Built-in
Ambient Temperature Compensation: With air flow of 1.1 SCFM (20 psig) maximum ambient is 204°C (400°F)
Dimensions: 9.55 L x 3.49 cm Dia. (3.76 x 1.375"); 0.9 m (36") PFA coated lead wires
Weight: 184 g (6.5 oz) with cable
**See range chart for linear ranges. Measurement is practical beyond thermocouple table values by use of radiation laws. Polynomials available upon request.