PX2780

Barometric Pressure Transducers with Terminal Strip

  • Environmental Monitoring Systems
  • Weather Measurement Systems
  • Environmental Data Logging
  • Laboratory and Test Station Monitoring
PX2780
From $ 981.50
Barometric Pressure Transducers with Terminal Strip PX2780 PX2780
loading all models
Product Overview
  • Burst Pressure 29 psi Maximum
  • Compensated Temperature, Max 60 °C
  • Compensated Temperature, Min -40 °C
  • Electrical Connection Removable Screw Terminal
  • Electrical Output 0 to 5 V
  • Media Compatibility Air, Argon, Carbon dioxide, Helium, Hydrogen, Nitrogen, Non-corrosive Dry Gasses, Oxygen
  • Mounting Mounting plate
  • Operating Temperature, Max 60 °C
  • Operating Temperature, Min -40 °C
  • Output Resistance <10 Ω
  • Output Signal Voltage
  • Process Connection Size 1/8 in
  • Process Connection Type Barbed Fitting
  • Proof Pressure Up to 1500 hpa
  • Resolution 0.1 hPa
  • Response Time <100 ms
  • Sensitivity Effect <± 0.09% full scale/50°C
  • Span Setting Adjustable Range ± 5 mV
  • Stability <± 0.1
  • Stability Note Long Term Stability: <± 0.1 hPa (mb)/year
  • Storage Temperature, Max 120 °C
  • Storage Temperature, Min -60 °C
  • Supply Power Note 9.5 to 28 Vdc (3 mA operate mode, 1 µA sleep mode) 3- or 4-wire circuit
  • Supply Voltage 9.5 to 28 Vdc
  • Thermal Effect on Span error, Max ± 1% gage
  • Warm-Up Shift <0.1 hPa (mb) maximum
  • Warm Up Time <1 sec
  • Weight 137 g
  • Zero/Balance Error <± 0.18% full scale/50°C
  • Zero Setting Adjustable Range ± 5 mV
The PX2780 high accuracy barometric pressure transducers are designed for applications in industry and science where accuracy, fast dynamic response and long term stability are important. The sensor and electronics are packaged in a rugged stainless steel and polyester case. A removable 5-pin terminal strip makes fast and secure connections. A high accuracy ceramic sensor is coupled with a digital compensation circuit using a custom ASIC to assure consistent accuracy and reliability over a broad range of environmental conditions. There is also a sleep mode where the unit only draws 1 micro amp that allows almost instant turn on/ reading for remote applications

SPECIFICATIONS
Accuracy: See table on next page
Resolution: 0.1 hPa (mb)
Long Term Stability: <± 0.1 hPa (mb)/year
Warm-Up Time: <1 second
Warm-Up Shift: <0.1 hPa (mb) maximum
Response Time: <100 ms
Proof Pressure: 1500 hPa (mb)
Burst Pressure: 2000 hPa (mb)
Storage Temperature: -60 to 120°C (-76 to 248°F)
Operating Temperature: -40 to 60°C (-40 to 140°F)
Compensated Range: -40 to 60°C (-40 to 140°F)
Zero Shift:
Barometric: <± 0.18% full scale/50°C (<± 0.2% full scale/100°F)
Sensitivity Effects: <± 0.09% full scale/50°C (<± 0.1% full scale/100°F)
Excitation: 9.5 to 28 Vdc (3 mA operate mode, 1 µA sleep mode) 3- or 4-wire circuit
Output: 0 to 5 Vdc
Full Scale Span Setting: ± 5 mV
Zero Setting: ± 5 mV
Output Impedance: <10 Ω
Electrical Connection: Removable 5-pin terminal strip
Pressure Connection: 1/8' barbed fitting
Pressure Media: Non-condensing air or gas
Weight: 137 g (4.8 oz)
Mounting: Mounting plate
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 5ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 24ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
en-US
en
Save Configuration