Rated 5 out of
5
by
sonoautomated from
Great little sensor.
We use these a lot to monitor surface temperatures. We like the quality and will continue to buy them.
Date published: 2016-02-25
Rated 4 out of
5
by
Pudge from
Sensor
We received this last week and it is functioning great.
Date published: 2018-04-25
- y_2025, m_4, d_30, h_7
- bvseo_bulk, prod_bvrr, vn_bulk_3.0.42
- cp_1, bvpage1
- co_hasreviews, tv_0, tr_2
- loc_en_CA, sid_SA2C-RTD-3-100-A-120, prod, sort_[SortEntry(order=HELPFULNESS, direction=DESCENDING), SortEntry(order=POSITIVE_FEEDBACK, direction=DESCENDING), SortEntry(order=SUBMISSION_TIME, direction=DESCENDING)]
- clientName_omega
- bvseo_sdk, java_sdk, bvseo-4.0.0
- CLOUD, getReviews, 5ms
- REVIEWS, PRODUCT
Hi Whats the accuracy of the sensor? Precision PT100 Class "B" or +-0,15C
Asked by: peter2
Product is Class A as defined by the DIN or IEC 60751 calibration standard. +/- 0.15°C at 0°C.
Answered by: NJENG
Date published: 2020-02-24
What is the response time on 60 degree C plate?
Asked by: seraph3
Thank you for your inquiry seraph3. Unfortunately, we do not have specific information about the response time of a probe as the environment will affect the response time directly. A time constant has been defined as the time required by a sensor to reach 63.2% of a step change in temperature under a specified set of conditions. Five time constants are required for the sensor to approach 100% of the step change value. An exposed junction thermocouple offers the fastest response. Also, the smaller the probe sheath diameter, the faster the response, but the maximum temperature may be lower. For more information, please review the following link: https://www.omega.co.uk/temperature/Z/ThermocoupleResponseTime.html
Answered by: Application Engineer MC
Date published: 2022-10-10
How does the SA2C compare to the SAPC models? For instance, SA2C-RTD-3-100-A-40 and SAPC-RTD-4-100-A-40 look to be identical.
Asked by: Evan12
The SAPC is a Class A element and the SA2C is Class B so a little less accurate. Physically pretty much identical and can be used interchangeably other then the Class of element.
Answered by: ApplicationEngineerANB
Date published: 2024-12-11
I'd like to know the weight of the sensor: SA2C-RTD-3-100-A-1M
Asked by: tisi
Thank you for your inquiry. We don't have this information readily available. Due to this being a more complicated issue please send an email to temp@omega.com for further assistance.
Answered by: Application Engineer MC
Date published: 2023-03-07
I want to use this sensor for running experiments on different surfaces at 300 Fdeg sunk in hot water. Can I do that with this sensor? how many times I can use this sensor (I plan to test 4 different surfaces).
Asked by: csrvivas
Unfortunately these probes are not for being immersed as are all other surface attached probes. The TFPS-TC series may be an option so check out it on www.omega.com.
Answered by: ApplicationEngineerANB
Date published: 2021-06-24
- y_2025, m_4, d_30, h_7CST
- bvseo_bulk, prod_bvqa, vn_bulk_3.0.42
- cp_1, bvpage1
- co_hasquestionsanswers, tq_5
- loc_en_CA, sid_SA2C-RTD-3-100-A-120, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
- clientName_omega
- bvseo_sdk, java_sdk, bvseo-4.0.0
- CLOUD, getContent, 9ms
- QUESTIONS, PRODUCT