OM-CP-XRTDTemp-Series
X-Series - Multi Channel RTD Temperature Loggers
Models In Stock
- ± 0.04 °C Calibrated Accuracy
- Real time Operation
- Up to 4 Hz Reading Rate
- Pushbutton or Programmable Start Time
OM-CP-XRTDTemp-Series
From
$
859.34
Models In Stock

Product Overview
- Calibration Digital calibration through software
- Data Interface WiFi
- Data Interface Note USB-A to micro USB cable (included); 460,800 baud
- Electrical Connection Screw Terminal
- Enclosure Aluminium
- Form factor Benchtop
- LED Indicators 1 per channel and 2 status LEDs
- Length 65.6 mm
- Operating Humidity 0 to 95% RH
- Operating Temperature, Max 60 °C
- Operating Temperature, Min -20 °C
- Process Temperature, Max 850 °C
- Process Temperature, Min -200 °C
- Process Temperature Note 18.5 Ω to 390.5 Ω -200 °C to +850 °C (-328 °F to +1562 °F) (Probe Dependent)
- Real Time Recording May be used with PC to monitor and record data in real time
- Samples,Max 2800000
- Sampling Rate 1 day
- Sampling Rate Note 4 Hz up to 1 reading every 24 hours
- Software Compatibility Windows XP SP3/7/8/10
- Start Modes Adjustable Delay Start, Immediate Start
- Supply Power Type Battery
- Temperature Resolution 0.001 °C
- Temperature Resolution Note 0.0001 Ω 0.001 °C (0.0018 °F) (Probe Dependent)
- Temperature Sensor Type RTD
- Time Accuracy ± 1 minute/month
- Width 184.2 mm
The new OM-CP-RTDTempX Series consists of 4, 8, 12 and 16-channel RTD Temperature data logger with a reading rate of up to 4 Hz. X-Series devices ship with a standard USB-A to Micro USB cable for use with free OM-CP PC based software. The X-series has a high memory capacity and increased download speed compared to previous generations of product.
To maximize memory capacity, users can enable or disable channels. For easy identification, each channel can be named with up to a ten digit title.
Using the OM-CP Data Logger Software, starting, stopping and downloading from the OM-CP-RTDTempX Series data loggers is simple and easy. Graphical, tabular and statistical data is provided for analysis and data can be viewed in multiple units, using the Engineering Units function. The data can also be automatically exported to Excel® for further calculations.
* Temperature specifications based on ideal 100 Ω Pt RTD compliant with IEC 751 (1983) and ITS-90, 5000 Ω FSR (accuracy based on 36 inch lead wire RTD with4 wire configuration)
** Calibrated accuracies based on standard Omega calibrations for 0 Ω to200 Ω range.
Specifications
TEMPERATURE | ||||
---|---|---|---|---|
Temprature | 18.5 Ω to 390.5 Ω | |||
Temprature | 0.0001 Ω | |||
Calibrated | ± 0.015 Ω @ 25 ℃ (77 ℉) ambient | |||
Specified | 18.5 Ω to 175.9 Ω | |||
Input | Removable screw terminal; | |||
Temprature | 2 ppm/℃ typical |
GENERAL | |||
---|---|---|---|
Memory | 4-channel: 698,709 readings per channel | ||
Start Modes | Software programmable immediate start | ||
Real Time | May be used with PC to monitor and | ||
LEDs | 1 per channel and 2 status LEDs | ||
Reading Rate | 4 Hz up to 1 reading every 24 hours | ||
Calibration | Digital calibration through software | ||
Calibration Date | Automatically recorded within device | ||
Battery Type | 9 V lithium included, user replaceable | ||
Battery Life | 18 months typical | ||
Data Format | Date and time stamped C,℉,K,ËšR;Ω µV | ||
Time Accuracy | ± 1 minute/month | ||
Computer | USB-A to micro USB cable (included); | ||
Operating | Windows XP SP3/7/8/10 | ||
Software | Standard Software version 4.2.19.0 or later | ||
Operating | -20 ℃ to +60 ℃(-4 ℉ to +140 ℉), | ||
Dimensions | 4-channel: 2.70 in × 7.25 in × 1.22 in | ||
Weight | 4-channel: 13 oz (368 g) | ||
Enclosure | Black anodized aluminum | ||
Approvals | CE, ROHS |
PDFs & Manuals
Software & Drivers
Show Ratings & Reviews
Could I read data directly to a .csv file via USB?
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@dwyeromega.com
Date published: 2024-11-19
Follow up to the question about reading temperature information on demand: can I read temperatures into LabVIEW in real time using the GUI software?
Thank you for your question. There is no Labview support for the logger just provided software.
Date published: 2021-12-03
Does it work Win10?
Thank you for your question. Yes it does work with Win 10.
Date published: 2021-04-29
Does it work under vacuum state? in the vacuum chamber?
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@OMEGA.COM
Date published: 2023-06-12
Can I put the entire logger in a vacuum chamber? I want to measure temperature in a vacuum enivroment
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@dwyeromega.com
Date published: 2025-01-28
Does it work with PT-1000 RTD's?
Thank you for your question. The logger only supports PT100 rtd's.
Date published: 2023-05-23
Why do channels lose connection when in use? is it because of ground loops with exposed thermocouples?
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@OMEGA.COM”
Date published: 2022-02-18
Is it possible to control the logger remotely using a computer? I would like to retrieve temperature information on demand, independent of the GUI software.
Thank you for your question. Unfortunately, the software is required to retrieve data from the logger via PC usb port.
Date published: 2021-11-03