1-800-663-4209 | Your feedback matters, take a quick survey Website Feedback

OM-CP-XVoltage-Series
X-Series - Multi Channel Voltage Loggers
Models In Stock
- Real-time Operation
- Programmable Engineering Units
- 24-bit Readings Provide High Resolution
- Up to 4 Hz Reading Rate
OM-CP-XVoltage-Series
From
$
862.77
Models In Stock

Product Overview
- Accuracy Detail Nominal range @ 25°C
- Additional Features 1 NIST Traceable Calibration
- Approvals CE, ROHS
- Baud Rate 460800
- Calibration Note Digital calibration through software, automatically recorded within device
- Conversion Time 150 milliseconds
- Current Measurement Range N/A
- Data Interface USB
- Electrical Connection Removable Screw Terminal
- Frequency Rejection 50 to 60 Hz
- Height 7.25 in
- LED Functionality 1 Per Channel and 2 Status LEDs
- Materials Aluminum
- Measurement Type Voltage
- Operating Temperature, Max 60 °C
- Operating Temperature, Min -20 °C
- Real Time Recording May be used with PC to monitor and record data in real time
- Relative Humidity 0 to 95% RH
- Sampling Rate Note 4 Hz to 1 reading every 24 hours
- Software Windows XP SP3/7/8/10, Standard Software version 4.2.19.0 or later, Secure Software version 4.2.18.0 or later
- Start Modes Button Push, Delayed Start, Immediate Start
- Supply Power 9 V Lithium Battery
- Supply Power Note 18 Months Typical
- Supply Power Type Battery
- Temperature Coefficient 25 ppm/°C
- Time Accuracy ±1 minute/month
- Width 2.7 in
The new OM-CP-VoltX Series consists of 4, 8, 12 and 16-channel, battery powered, DC voltage data logger with a reading rate of up to 4 Hz. X-Series devices ship with a standard USB-A to Micro USB cable for use with free OMCP PC based software. The X-series has a high memory capacity and increased download speed compared to previous generations of product.
This easy-to-use device features an 18 month typical battery life, user replaceable battery and programmable engineering units. The storage medium is non-volatile solid state memory, providing maximum data security even if the battery becomes discharged. To maximize memory capacity, users can enable or disable channels. For easy identification, each channel can be named with up to a ten digit title.
Using the OM-CP Data Logger Software, starting, stopping and downloading from the OM-CP-VoltX Series data loggers are simple and easy. Graphical, tabular and statistical data is provided for analysis and data can be viewed in multiple units, using the Engineering Units function. The data can also be automatically exported to Excel® for further calculations.
Specifications
MEASUREMENT | ||||
---|---|---|---|---|
Input | Removable screw terminals | |||
Model | 160 mV | 3.2 V | 32 V | |
Voltage Range | ±160 mV | ±3.2 V | -8 V to +32 V | |
Voltage | 0.02 mV | 0.0004 mV | 0.004 mV | |
Calibrated | ±0.032 mV | ±0.0032 V | ±0.02 V | |
Input Impedance | 1 MΩ | 125 kΩ | 125 kΩ | |
Maximum | +3 V | +50 V | +50 V | |
Analog | 150 ms | |||
Frequency | 60+50 Hz | |||
Temprature | 25 ppm/ °C | |||
Specified | Nominal range @ 25 °C | |||
Engineering | Native Measurement units can be scaled |
GENERAL | |||
---|---|---|---|
Memory | 4-channel: 698,709 readings per channel | ||
Start Modes | Software programmable immediate start | ||
Real Time | May be used with PC to monitor and | ||
LEDs | 1 per channel and 2 status LEDs | ||
Reading Rate | 4 Hz up to 1 reading every 24 hours | ||
Calibration | Digital calibration through software | ||
Calibration Date | Automatically recorded within device | ||
Battery Type | 9 V lithium included, user replaceable | ||
Battery Life | 18 months typical | ||
Data Format | Date and time stamped V, mV, µV | ||
Time Accuracy | ±1 minute/month | ||
Computer | USB-A to micro USB cable (included); | ||
Operating | Windows XP SP3/7/8/10 | ||
Software | Standard Software version 4.2.19.0 or later | ||
Operating | -20 °C to +60 °C(-4 °F to +140 °F), | ||
Dimensions | 4-channel: 2.70 in × 7.25 in × 1.22 in | ||
Weight | 4-channel: 13 oz (368 g) | ||
Enclosure | Black anodized aluminum | ||
Approvals | CE, ROHS |
PDFs & Manuals
Software & Drivers
Show Ratings & Reviews
How do I get the data logger to sample at 4hz? The drop-down menu's fastest sample rate is 1/sec.
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@OMEGA.COM
Date published: 2022-04-28