1-800-663-4209 | Welcome to our new website, found an issue or bug? Please report it here
868-869
1 Channel 3 & 4 Wire PT100 RTD Temperature Meter
- Each Unit Comes with Certificate Stating NIST Traceability and ANSI/Z540-1-1994 Compliance
- 3- and 4-wire RTD’s
- Accurate
- 0.1° and 1° Resolution

Product Overview
- Battery life 500 hr
- Display Note Display Units : °C or °F
- Display Type Note 3 1⁄2 digit LCD, 13 mm (0.5") height; polarity, decimal point, open sensor and overrange indication
- Height 160 mm
- Lead Resistance, Max 4-wire: 50 Ω max; 3-wire: 10 Ω max
- Length 31 mm
- Measuring Temperature Range, Max 630 °C
- Measuring Temperature Range, Min -100 °C
- Operating Temperature, Max 50 °C
- Operating Temperature, Min -10 °C
- Operating Temperature Note Operating: -10 to 50°C (14 to 122°F), less than 80% RH, up to 35°C
- Relative Humidity, Max 0.9
- Relative Humidity, Min 0%
- Repeatability <0.2 °F
- Sampling Rate 1.5 Hz (1.5 updates/second)
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -35 °C
- Storage Temperature Note -35 to 60°C (-31 to 140°F), less than 90% RH, up to 35°C
- Supply Power Note Maximum Common Mode Voltage: 42V peak
- Supply Voltage 9 Vdc
- Temperature Accuracy ± 0.3 °C
- Temperature Accuracy Note 200°C: Accuracy: ± 0.3°C; ± 1.5°C 1100°C: Accuracy: ± 1°C; ± 2°C Includes DIN 43760 conformity (no points), repeatability, temperature coefficient (18 to 28°C ambient), time stability (1 year) and errors with up to 50 Ω of resistance in each lead; excluded probe errors
- Temperature Coefficient 18 to 28°C (64 to 82°F); included in accuracy specification; less than 18°C (64°F) or greater than 28°C (82°F) less than ± (0.015°C)/°C
- Temperature Resolution 0.1 °C
- Temperature Resolution Note 200°C: Resolution: 0.1°C 1100°C: Resolution: 1°C
- Temperature Sensor Type RTD
- Temperature Sensor Type Note 3- or 4-wire 100 Ω Platinum RTD a=0.00385
- Weight 210 g
- Width 69 mm
SPECIFICATIONS
Sensor: 3- or 4-wire 100 Ω Platinum RTD α=0.00385
Accuracy: Includes DIN 43760 conformity (no points), repeatability, temperature coefficient (18 to 28°C ambient), time stability (1 year) and errors with up to 50 Ω of resistance in each lead; excluded probe errors
Repeatability: Less than 0.2°F typical for one week at constant ambient temperature
Temperature Coefficient: 18 to 28°C (64 to 82°F); included in accuracy specification; less than 18°C (64°F) or greater than 28°C (82°F) less than ± (0.015°C)/°C
Maximum Lead Resistance: 4-wire:50 Ω max; 3-wire: 10 Ω max
Display: 3½ digit LCD, 13 mm (0.5") height; polarity, decimal point, open sensor and overrange indication
Reading Rate: 1.5 per second
Battery Life: 500 hours typical, with 9V alkaline battery (included)
Battery Indicator: Display reads "LO BAT" when less than 10% of life remains
Linearization: Ratiometric dual-slope A/D with continuous linearization
Mating Connection: OMEGA series T instrumentation connectors (model TA4F and TA4FL), see section G
Environmental Limits: Operating: -10 to 50°C (14 to 122°F), less than 80% RH, up to 35°C
Storage: -35 to 60°C (-31 to 140°F), less than 90% RH, up to 35°C
Maximum Common Mode Voltage: 42V peak
Dimensions: 160 H x 69 W x 31 D mm(6.3 x 2.7 x 1.2")