Product Overview
- Accuracy ± 0.025 %
- Accuracy Note ± (0.025% of reading + 0.05 Ω)
- Additional Features 1 Read & Source Signals
- Battery Indication A low battery indication is displayed when the battery voltage drops below the operating level
- Battery life 50 hr
- Display Type LCD
- Display Type Note High contrast graphic liquid crystal display, LED backlighting for use in low lit areas
- Height 40.64 mm
- Length 143 mm
- Operating Temperature, Min -25 °C
- Operating Temperature Note Nominal Temperature: 23°C
- Protection Over-voltage protection to 60 V dc (rated for 30 seconds) (Protection against misconnection)
- Relative Humidity, Max 0.9
- Relative Humidity, Min 10%
- Relative Humidity Note 10% ≤RH ≤90% (0 to 35°C), non-condensing 10% ≤RH≤ 70% (35 to 60°C), non-condensing, Nominal Relative Humidity: 70%
- Supply Voltage 1.5 Vdc
- Temperature Drift ± 0.01%
- Temperature Drift Note ± 0.01% of span outside of 23°C ± 10°C (73°C ± 18°F
- Thermocouple Type None
- Type of RTD Signal 100 Ω, 1000 Ω
- Voltage Read Excitation Current: 0.5 mA nominal Normal Mode Rejection: 50/60 Hz, 50 dB Common Mode Rejection: 50/60 Hz, 120 dB
- Voltage Source Accuracy From 1 to 10.2 mA External Excitation Current: ± (0.025% of Reading + 0.05 Ω) Below 1 mA of External Excitation Current: ± (0.025% of Reading + 0.025 mV/ mA Excitation Current + 0.05 Ω) Resistance Ranges: 0.00 to 410.00, 410.1 to 2001.0 Ω Allowable Excitation Current Range: <410 Ω: 10.2 mA max; steady or pulsed/intermittent 410 to 2001 Ω: 1 mA max; steady or pulsed/intermittent Pulsed Excitation Current Compatibility: DC to 0.01 second pulse width
- Weight 12.1 oz
- Weight Note 12.1 oz (including boot and batteries)
- Width 76.2 mm
With the CL514-PLUS you can check and calibrate all your RTD instruments and measure RTD sensors. Automatic indication of connections on the display for simple hookups. Take it without worry into the shop, plant or field. It comes protected with a rubber boot and rugged, low profile switch. Easy to operate even in the dark areas of the plant with the backlit display. Stop carrying around a decade box and RTD resistance tables, you can calibrate directly in temperature (°C and °F). The CL514-PLUS works with a wide variety of RTD curves including Platinm 100 (alpha = 3850, 3902, 3926) and 1000 (alpha = 3850, 3750) ?, Copper 10 and 50 ?, Nickel 100 and 120 ?. Easily set any value quickly to within 0.1° with the adjustable digital potentiometer "DIAL" plus store any three temperatures for instant recall. Calibrate quickly with automatic output stepping. Choose between 2, 3, 5, 11 and 21 steps to automatically increment the output in 100%, 50%, 25%, 10% or 5% of span. Select the step time to match your system from 5, 6, 7, 8, 10, 15, 20, 25, 30 and 60 seconds. Connect directly to RTD inputs of smart transmitters, PLCs, controllers and multichannel recorders and verify their outputs or displays. Measure and trouble shoot RTD sensor connections and find broken wires. Connect your 2, 3 or 4 wire RTD sensor and the CL541-PLUS automatically measures the RTD in °C or °F.
SPECIFICATIONS (Unless otherwise indicated all specifications are rated from a nominal 23°C, 70% RH)
Accuracy: ± (0.025% of Reading + 0.05 ?)
Temperature Drift: ± 0.01% of span outside of 23°C ± 10°C (73°C ± 18°F)
Operating Temperature Range: -25 to 60°C (-10 to 140°F)
Relative Humidity Range: 10% ?RH ?90% (0 to 35°C), non-condensing 10% ?RH ?70% (35 to 60°C), non-condensing
Dimensions: 143 L x 76.2 W x 40.64 mm H (5.63 x 3.00 x 1.60")
Weight: 12.1 oz (including boot and batteries)
Batteries: 4 "AA" Alkaline 1.5V (included)
Battery Life: 50 Hours
Low Battery: Low battery indication with nominal 1 hour of operation left
Protection Against Misconnection: Over-voltage protection to 60 V dc (rated for 30 seconds)
Display: High contrast graphic liquid crystal display. LED backlighting for use in low lit areas.
READ
Excitation Current: 0.5 mA nominal
Normal Mode Rejection: 50/60 Hz, 50 dB
Common Mode Rejection: 50/60 Hz, 120 dB
Source
Accuracy From 1 to 10.2 mA External Excitation Current: ± (0.025% of Reading + 0.05 ?) Below 1 mA of External Excitation
Current: ± (0.025% of Reading + 0.025 mV/ mA Excitation Current + 0.05 ?)
Resistance Ranges: 0.00 to 410.00, 410.1 to 2001.0 ? Allowable Excitation Current Range:
<410 ?: 10.2 mA max; steady or pulsed/intermittent
410 to 2001 ?: 1 mA max; steady or pulsed/intermittent Pulsed Excitation Current
Compatibility: DC to 0.01 second pulse width
SPECIFICATIONS (Unless otherwise indicated all specifications are rated from a nominal 23°C, 70% RH)
Accuracy: ± (0.025% of Reading + 0.05 ?)
Temperature Drift: ± 0.01% of span outside of 23°C ± 10°C (73°C ± 18°F)
Operating Temperature Range: -25 to 60°C (-10 to 140°F)
Relative Humidity Range: 10% ?RH ?90% (0 to 35°C), non-condensing 10% ?RH ?70% (35 to 60°C), non-condensing
Dimensions: 143 L x 76.2 W x 40.64 mm H (5.63 x 3.00 x 1.60")
Weight: 12.1 oz (including boot and batteries)
Batteries: 4 "AA" Alkaline 1.5V (included)
Battery Life: 50 Hours
Low Battery: Low battery indication with nominal 1 hour of operation left
Protection Against Misconnection: Over-voltage protection to 60 V dc (rated for 30 seconds)
Display: High contrast graphic liquid crystal display. LED backlighting for use in low lit areas.
READ
Excitation Current: 0.5 mA nominal
Normal Mode Rejection: 50/60 Hz, 50 dB
Common Mode Rejection: 50/60 Hz, 120 dB
Source
Accuracy From 1 to 10.2 mA External Excitation Current: ± (0.025% of Reading + 0.05 ?) Below 1 mA of External Excitation
Current: ± (0.025% of Reading + 0.025 mV/ mA Excitation Current + 0.05 ?)
Resistance Ranges: 0.00 to 410.00, 410.1 to 2001.0 ? Allowable Excitation Current Range:
<410 ?: 10.2 mA max; steady or pulsed/intermittent
410 to 2001 ?: 1 mA max; steady or pulsed/intermittent Pulsed Excitation Current
Compatibility: DC to 0.01 second pulse width