BB702-series
Ambient 10 °C to 215 °C Temperature Blackbody Calibrator
- Model BB702 Calibrates from Ambient +10 to 215°C (Ambient +20 to 420°F)
- Portable Rugged Design
- Calibrates Infrared Pyrometers Quickly and Accurately
- Built in Digital PID Autotune Temperature Controller with Temperature Readout
BB702-series
From
C$
5 026.13

Product Overview
- Accuracy ± 0.5 °C
- Accuracy Note ±0.5°C, ±0.25% rdg (±0.9°F, ±0.25% rdg)
- Additional Features 1 Portable
- Calibration Reference Sensor Platinum RTD, class A, alpha = 0.00385
- Cavity Emissivity 0.95
- Control Mode PID Control
- Data Interface RS232
- Display Type LED
- Frequency 50/60 Hz
- Height 152 mm
- Installation Category II
- Internal Sensing Element Platinum RTD, class A, alpha = 0.00385
- Length 280 mm
- Operating Temperature, Max 45 °C
- Operating Temperature, Min 5 °C
- Power Consumption 75 W
- Process Temperature, Max 215 °C
- Process Temperature, Min 10 °C
- Stability ±0.1°C
- Target Size 63.5
- Warm Up Time 12 min
- Warm-Up Time Note 12 min from ambient to 204°C (400°F)
- Width 305 mm
The BB702 Blackbody Calibrator is a high performance, rugged, portable calibrator for infrared pyrometers. The BB702 has a range from Ambient +10 to 215°C (Ambient +20 to 420°F) with an accuracy of ±0.5°C, ±0.25% rdg. The unit provides a heated target plate of known temperature and emissivity. Virtually any infrared pyrometer with a spot size diameter of 63.5 mm (2.5") or smaller can be calibrated. Both models come with an RS-232 computer interface which allows computer control of the setpoints for automatic test applications. An NIST traceable calibration certificate is also provided.
PLEASE NOTE: Items listed as "Stock" require 5 Business Days for NIST Certification and Order Processing
PLEASE NOTE: Items listed as "Stock" require 5 Business Days for NIST Certification and Order Processing
PDFs & Manuals
Show Ratings & Reviews
What is the uncertainty of the target plate's 0.95 emissivity? Or is that uncertainty already incorporated into the plate's temperature accuracy?
Hi Hartman,
Thank you for the inquiry. Accuracy of the emissivity is included in the calibrator accuracy.
Best regards
OMEGA
Date published: 2023-09-07