Welcome to our new website, found an issue or bug? Please report it here

RH318-Series
Handheld Hygro-Thermometer Data Logger
- Thermistor Sensor for Temperature Measurement, Fast Response Time
- Large Backlit Dual LCD Display, MAX/MIN/HOLD Function
- Dew Point Measurement
- 16,000 Record Data Logger
RH318-Series
From
C$
565.89

Product Overview
- Additional Features 1 Data Logging Capability
- Battery life 80 hr
- Height 185 mm
- Humidity Response Time 180 sec
- Hysteresis <1.5%RH
- Length 65 mm
- Measurement Type Dew Point, Relative Humidity, Temperature
- Measuring Temperature Range, Max 60 °C
- Measuring Temperature Range, Min -20 °C
- Probe Diameter 15 mm
- Probe Length 94 mm
- Process Temperature, Max 60 °C
- Process Temperature, Min -20 °C
- Response Time 10 sec
- Sampling Rate 2 Hz (2 updates/second)
- Sampling Rate Note 2 Times/sec
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -10 °C
- Supply Voltage 9 Vdc
- Temperature Accuracy ± 0.8 °C
- Temperature Accuracy Note Probe: RP-32 90%RH ± 4.0%RH 10%RH to 90%RH ± 2.0%RH -20 to 60°C (-4 to 140°F) ± 0.8°C (± 1.5°F) Probe: RP-31 90%RH ± 5.0%RH 10%RH to 90%RH ± 3.0%RH -15 to 35°C (-60 to 95°F) ± 0.8°C (± 1.5°F) 35°C (95°F)
- Temperature Coefficient 0.1 x (specified accuracy) per 1°C (<23°C or >23°C)
- Temperature Resolution 0.1 °C
- Temperature Resolution Note 0.1%RH, 0.1°C, 0.1°F
- Temperature Sensor Type Capacitance
- Weight 285 g
- Weight Note Approximately 285 g (10 oz)
- Width 36 mm
The RH318 is a low cost, high performance handheld meter with data logging ability. It offers a large dual back-lit display to show the temperature and humidity from a handheld probe and can display a dew point measurement. The temperature/humidity probe uses a semiconductor and polymer capacitive sensor and a thermistor temperature sensor. Windows® based software and USB interface cable is included to allow real time or recorded data to be read by a PC through the USB/RS232 port (RS232 interface cable is optional). Probe can extend up to 1.5 m (5') with optional extension cable.
SPECIFICATIONS
Sensor Type: Electronic capacitance polymer film sensor/NTC
Relative Humidity Range: 0 to 100% RH
Temperature Range: -20 to 60°C (-4 to 140°F)
Dew Point Temperature Range: -20 to 60°C (-4 to 140°F)
Resolution: 0.1% RH, 0.1°C, 0.1°F
Accuracy Probe (RH318-RP): <10% RH, >90% RH ± 4.0% RH 10% RH to 90% RH ± 2.0% RH -20 to 60°C (-4 to 140°F) ± 0.8°C (± 1.5°F)
Response Time (In Slow Moving Air):
Humidity: 180 s
Temperature: 10 s
Sensor Hysteresis: <1.5% RH
Temperature Coefficient: 0.1 x (specified accuracy) per 1°C (<23°C or >23°C)
Sample Rate: Two times per second
Battery Type: 9V battery (included)
Battery Lifetime: Approx. 80 hours (alkaline battery)
AC Adaptor (Optional): 9 Vdc (8 to 10 Vdc max )
Operation Temperature: 0 to 40°C (32 to 104°F)
Operation Humidity: 10 to 90% RH
Storage Temperature: -10 to 60°C (14 to 140°F)
Storage Humidity: 10 to 75% RH
Dimensions
Meter: 185 L à— 65 W à— 36 mm H (7.3 x 2.6 x 1.4")
Probe: 15 D x 94 mm L (0.6 x 3.7")
Weight: Approx. 285 g (10 oz)
Environment Conditions:
Altitude: Up to 2000 m (6562')
Relative Humidity: 90% max
Operation Ambient Temperature: 0 to 40°C (32 to 104°F)
SPECIFICATIONS
Sensor Type: Electronic capacitance polymer film sensor/NTC
Relative Humidity Range: 0 to 100% RH
Temperature Range: -20 to 60°C (-4 to 140°F)
Dew Point Temperature Range: -20 to 60°C (-4 to 140°F)
Resolution: 0.1% RH, 0.1°C, 0.1°F
Accuracy Probe (RH318-RP): <10% RH, >90% RH ± 4.0% RH 10% RH to 90% RH ± 2.0% RH -20 to 60°C (-4 to 140°F) ± 0.8°C (± 1.5°F)
Response Time (In Slow Moving Air):
Humidity: 180 s
Temperature: 10 s
Sensor Hysteresis: <1.5% RH
Temperature Coefficient: 0.1 x (specified accuracy) per 1°C (<23°C or >23°C)
Sample Rate: Two times per second
Battery Type: 9V battery (included)
Battery Lifetime: Approx. 80 hours (alkaline battery)
AC Adaptor (Optional): 9 Vdc (8 to 10 Vdc max )
Operation Temperature: 0 to 40°C (32 to 104°F)
Operation Humidity: 10 to 90% RH
Storage Temperature: -10 to 60°C (14 to 140°F)
Storage Humidity: 10 to 75% RH
Dimensions
Meter: 185 L à— 65 W à— 36 mm H (7.3 x 2.6 x 1.4")
Probe: 15 D x 94 mm L (0.6 x 3.7")
Weight: Approx. 285 g (10 oz)
Environment Conditions:
Altitude: Up to 2000 m (6562')
Relative Humidity: 90% max
Operation Ambient Temperature: 0 to 40°C (32 to 104°F)
PDFs & Manuals
Show Ratings & Reviews
I'm looking for a calibration (adjustment) procedure for the Omega RH318 unit - Would you be willing to provide that to me, as I believe our company (Transcat) is a distributor for your products?
Hi John78
Thank you for the inquiry. Calibration procedure is available.Please forward request to temp@omega.com.
Best regards
OMEGA
Date published: 2023-02-01