Skip to search Skip to main content Skip to navigation

Welcome to our new website, found an issue or bug? Please report it here

We noticed you are from . Visit your local site for regional offers and live support.

LCEB

Low Profile, Mini Beam Load Cell for Low Capacities

Models In Stock
  • Very High Precision
  • Low Profile - 1" Height
  • High Repeatability
  • 100% Creep Tested
LCEB
From C$ 394.76
Models In Stock
Low Profile, Mini Beam Load Cell for Low Capacities LCEB LCEB
loading all models
Product Overview
  • Cable Length 5 ft
  • Calibration Option NIST
  • Creep in 20 min ± 0.15%
  • Electrical Connection Cable
  • Hysteresis ± 0.02%
  • Input Resistance 350 +50/-3.5 Ω
  • Load Cell Type Bending Beam
  • Operating Temperature, Max 90 °C
  • Operating Temperature, Min -55 °C
  • Output Resistance 350 ± 3.5 Ω
  • Overload, Max ± 400% FS
  • Process Temperature, Max 65 °C
  • Process Temperature, Min -15 °C
  • Repeatability ± 0.01%
  • Safe Overload 150% FS
  • Supply Power Note 10 Vdc, 15 Vdc maximum
  • Supply Voltage 10 to 15 Vdc
  • Thermal Span Effect ± 0.0008% FS/°F
  • Thermal Zero Effect ± 0.0015% rdg/°F
  • Wiring/Connector Table Green: (+) Output; White: (-) Output; Black: (-) Input; Red: (+) Input
  • Zero/Balance Error ± 1%
Model LCEB load cells offer exceptional accuracy in compression in controlled environments (in general anywhere a digital readout is used). Their high accuracy, small size, low creep, and excellent temperature compensation make the LCEB series an excellent choice for applications such as force testing, thrust measurements, check weighing, and scale applications.
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 3ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;

Hi, Could we use 5vdc excitation with this product? Does it affect the precision or the linearity spec ? Thanks

Asked by: Mat27
You can use 5VDC without effecting the accuracy or linearity. The only difference would be that the output signal of the load cell would be 0 to 15mV instead of 0 to 30mV
Answered by: Applications Engineer JK
Date published: 2023-06-09

Would this LCEB series load cell hold up to tests with very high relative humidity? (10-100%) Would it's measuring capabilities be compromised?

Asked by: cm10w13pts
The LCEB has an IP65 environmental rating. It should be able to handle dust and water ingress
Answered by: JK Applications Engineer
Date published: 2024-02-19
  • y_2025, m_2, d_20, h_9CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.42
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_2
  • loc_en_CA, sid_LCEB-10, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 29ms
  • QUESTIONS, PRODUCT
en-CA
en_CA
Save Configuration