Welcome to our new website, found an issue or bug? Please report it here

DPG409S
Sanitary, High Accuracy, Digital Pressure Gauge with Output
- Easy to Read 25 mm (1') Backlit LCD Display
- High 0.08% Accuracy Typical
- Hi and Lo Alarms Standard
- MIN/MAX/AVG Reading
DPG409S
From
C$
1,632.33

Product Overview
- Accuracy Class < 0.1%
- Accuracy Note ± 0.08% Best Straight Line Typ. (Linearity, Hysteresis and Repeatability combined)
- Battery life 9 mo
- Battery Size and Type 3.6V Li-ion
- Burst Pressure 6x Full Scale
- Cable Length 2.7 m
- Calibration Note 5-Pt NIST traceable Calibration with zero and span values, Calibrated in horizontal direction (Compound Gage models calibrated in positive direction only)
- Compliance CE
- Connection Location Lower
- Data Interface USB
- Data Interface Note USB connection for set-up, cable included
- Display Segment Height 25 mm
- Display Type 4 Digit LCD
- Display Type Note 4-digit LCD with user selectable backlighting, (9999 count) 25 mm (1') character height.
- Electrical Output 4 to 20 mA
- Electrical Output Note User selectable 0 to 5 Vdc, 0 to 10 Vdc of 4 to 20 mA (external 24 Vdc power required)
- Enclosure Material 316 Stainless Steel
- Enclosure Ratings IP65, NEMA Type 4X, Weatherproof
- Fill Fluid NEOBEE M5 food grade
- Isolation between Case and Output Terminations, Min 100MΩ @ 50 Vdc
- Materials, Wetted 316 Stainless Steel
- Number of Batteries 1
- Number of Digits 4
- Operating Temperature, Max 66 °C
- Operating Temperature, Min -18 °C
- Output Signal Current, Voltage
- Pressure Accuracy ± 0.08 %
- Pressure Cycles 250000
- Process Connection Note 11 ⁄2 and 2" Tri-GripTM Fitting (compatible with Tri-Clamp®): 600 psi using suitable clamp (600 psi on 11 ⁄2" Tri-Clamp® only)
- Process Connection Type Tri-Clamp
- Sample/Display Rate 0.25 to 10 sec
- Sample Display Rate Note User selectable from 0.25 to 10 sec. (preset to 1/sec)
- Sensor Dimension Note 72 to 89 mm L (2.8 to 3.5") below display
- Settable Parameters Backlight, Min, Max, Zero, Clear
- Software Settings Software Settings (via USB) Update Rate: 4 options, 10, 5, 2, 1 times per second Units: psi, inHg, inH2O, bar, mbar, hPa units; must be specified at time of order Lock: Allows the lock-out of front panel functions Alarms: User selectable “High” and “Low” alarm limits (open collector) Backlight: On/off, 10 sec, 30 sec, 1 min, 5 min Wireless Transmitter (Optional): Channel number, transmission rate, alarms, sensor offset, chart recording, data logging Calibrate: Zero and Span Overrange Indication: “_ _ _ _”
- Span Setting ± 0.5% FS typical ± 1% maximum (For ranges ≤2.5 psi ± 1% Typical ± 2% maximum)
- Storage Temperature, Max 82 °C
- Storage Temperature, Min -40 °C
- Supply Power Type Battery
- Thermal Effects Thermal Effect (Over Compensated Range): Zero and Span shift over Compensated Range; Ranges: ≤2.5 psi: Zero: 1.00% Span: 0.70% ;5 psi: Zero: 0.80% Span: 0.50% ; > 5 psi: Zero: 0.50% Span: 0.50% ; 26B & 26HB Barometric Range: Zero: 0.90% Span: 0.50%
- Weight 900 g
- Weight Note 900 g (2 lb) Typical depending upon configuration
- Wireless Option Specifications Transmit Sample Rate: User programmable from 1 sample/2 min to 1 sample/2 sec Radio Frequency (RF) Transceiver Carrier: ISM 2.4 GHz RF Output: 10 dBm (10 mW) Range of RF Link: Outdoors (Line of Sight): Up to 120 m (400') Indoors/Urban: Up to 40 m (130') Included Software: Requires Windows® 7 (32 bit); XP, Vista Internal Battery: Two 3.6V lithium 8.4 Ah capacity (C-cell) included Data Transmission to Host: Pressure reading, ambient temperature reading, RF transmit strength and battery level
- Zero/Balance Error ± 0.5% FS typical ± 1% maximum (For ranges ≤2.5 psi ± 1% Typical ± 2% maximum)
SPECIFICATIONS
RoHS Compliant
CE Compliant: Meets industrial emissions and immunity standard IEC61326 for industrial locations
Accuracy: ± 0.08% Best Straight Line Typ. (Linearity, Hysteresis and Repeatability combined)
Calibration: 5-Pt NIST traceable Calibration with zero and span values, Calibrated in horizontal direction (Compound Gage models calibrated in positive direction only).
Zero Balance: ± 0.5% FS typical ± 1% maximum (For ranges =2.5 psi ± 1% Typical ± 2% maximum)
Span Setting: ± 0.5% FS typical ± 1% maximum (For ranges =2.5 psi ± 1% Typical ± 2% maximum)
Storage/Operating Temperature Range:
Storage: -40 to 82°C (-40 to 180°F)
Operating: -18 to 66°C (0 to 150°F)
Compensated Temperature Range:
Ranges =5 psi: 4.4 to 60°C (40 to 140°F)
Ranges >5 psi: -4 to 85°C (25 to 150°F)
Thermal Effect (Over Compensated Range):
Zero and Span shift over Compensated Range
Ranges =2.5 psi: Zero: 1.00%
Span: 0.70%
Range = 5 psi: Zero: 0.80%
Span: 0.50%
Ranges > 5 psi: Zero: 0.50%
Span: 0.50%
26B & 26HB Barometric Range:
Zero: 0.90% Span: 0.50%
Minimum Isolation Between Case and Output Terminations: 100M O @ 50 Vdc
Analog Output: User selectable 0 to 5 Vdc, 0 to 10 Vdc of 4 to 20 mA (external 24 Vdc power supply required for all analog output types)
Note: Compound gage ranges have a unidirectional analog output equal to their full span; i.e. -15 to +15 psi = 0 to 10 Vdc.
Analog Output Cable: 2.7 m (8.9')
Analog Output Accuracy: 0.08% typical
Pressure Cycles: 250,000, minimum
Long Term Stability (1-Year): ± 0.1% FS Typ.
Fill Fluid: NEOBEE M5 Food Grade
Approved to 3A Sanitary Specification
Environmental Protection: NEMA 4X (IP65) depending upon electrical termination
Pressure Rating: 1½ and 2' Tri-GripTM Fitting (compatible with Tri-Clamp®): 600 psi using suitable clamp (600 psi on 1½' Tri-Clamp®) only)
Over Pressure: 4 x rated pressure to maximum rated flange pressure
Secondary Containment: 6 x rated pressure to a maximum of 2000 psi (not in clamp)
Display: 4-digit LCD with user selectable backlighting, (9999 count) 25 mm (1') character height
Computer Interface: UBS connection for set-up, cable included
Sample/Display Rate: User selectable from 0.25 to 10 sec. (preset to 1/sec)
Power: One 3.6V lithium 8.4 Ah capacity (C cell) Included (Two included with Wireless Option)
Battery Life (Typical): Up to 9 months
Enclosure Material: 316 Stainless Steel and ABS center gasket
Enclosure Finish: Brushed 316 stainless steel
Enclosure Environmental Rating: Weatherproof, NEMA 4X (IP65)
Connection Location: Lower (head rotates to position display)
Front Panel Functions
Backlight: Manually turns backlight on/off
Min/Max: Recalls min/max values
Zero/Clear: Zeros display or clears min/max
Software Settings (via USB) Update Rate: 4 options, 10, 5, 2, 1 times per second
Units: psi, inHg, inH2O, bar, mbar, hPa, units must be specified at time of order
Lock: Allows the lock-out of front panel functions
Alarms: User selectable “High” and “Low” alarm limits. (Open Collector)
Analog Output: User Selectable 0 to 5 Vdc, 0 to 10 Vdc, or 4 to 20 mA (requires external 24 Vdc power)
Backlight: On/Off, 10 sec, 30 sec, 1 min, 5 min
Wireless Transmitter (optional): Channel Number, Transmission rate, alarms, sensor offset, chart recording, data logging
Calibrate: Zero and Span
Overrange Indication: “_ _ _ _”
Wireless Option Sepcifications
Transmit Sample Rate: User programmable from 1 sample/2 min to 1 sample/2 sec
Radio Frequency (RF) Transceiver Carrier: ISM 2.4 GHz
RF Output: 10 dBm (10 mW)
Range of RF Link:
Outdoors (Line of Sight): Up to 120 m (400')
Indoors/Urban: Up to 40 m (130')
Included Software: Requires Windows® 7 (32 bit); XP, Vista
Internal Battery: Two 3.6V lithium 8.4 Ah capacity (C-cell) included
Data Transmission to Host: Pressure reading, ambient temperature reading, RF transmit strength and battery level
Dimensions:
Display: 115 D x 89 W x 74 mm H (4.5 x 3.5 x 2.9')
Sensor: 72 to 89 mm L (2.8 to 3.5') below display
Weight: 900 g (2 lb) Typical depending upon configuration
PDFs & Manuals
Software & Drivers
Show Ratings & Reviews
What is the wire connection or wire size of Wire Gland/Compression Fitting? Manual shows 5 screw terminals inside, and mentions "Analog Output Cable (TX4)", but this appears to not correspond to any wire thickness standard. Which connector or wire to use?
You should use a 4 conductor shielded cable. I the individual lead size should be 24 AWG. Overall O.D. of the cable should be 0.20"
Date published: 2024-03-05
Is there an audible signal when the high or low alarm is triggered?
This unit does not come with inaudible signal for the alarms, you can add an audible alarm. Please see manual for instructions
Date published: 2021-05-12