Welcome to our new website, found an issue or bug? Please report it here

lcm101-lcm111
Metric Stainless Steel S-Beam Load Cells
Models In Stock
- 5 Point NIST Traceable Calibration with
59 kOhm Shunt Cal Data - All Stainless Steel for Harsh Industrial Applications
- 0.25% Interchangeable for Multiple Load Cell Applications
lcm101-lcm111
From
C$
766.84
Models In Stock

Product Overview
- Calibration Option 5 POINT NIST
- Electrical Output 3 mV/V
- Force Type Compression, Tension
- Full Scale Deflection 0.02 in
- Hysteresis ± 0.02%
- Input Resistance 350 ± 10 Ω
- Interchangeability 0.0025
- Linearity ± 0.03%
- Load Cell Type S-beam
- Materials 17-4 PH Stainless Steel
- Operating Temperature, Max 93 °C
- Operating Temperature, Min -40 °C
- Output Resistance 350 ± 10 Ω
- Overload, Max 300% FS
- Process Temperature, Max 71 °C
- Process Temperature, Min 17 °C
- Repeatability ± 0.01%
- Safe Overload 150% FS
- Supply Power Note 10 Vdc, 15 Vdc maximum
- Supply Voltage 10 Vdc
- Thermal Span Effect 0.0018% FSO/°C
- Thermal Zero Effect 0.0018% FSO/°C
- Zero/Balance Error ± 1%
The LCM101 & LCM111 are also available in standard configurations, with 0-25 to 0-40,000 lb ranges and standard threads. Please see the LC101 for complete details, or contact our Pressure department for more information.
SPECIFICATIONS
Excitation: 10 Vdc, 15 Vdc max
Output: 3 mV/V ± 0.0075 mV/V
Linearity: ± 0.03% FSO (0.1% 40 K)
Hysteresis: ± 0.02% FSO (0.1% 40 K)
Repeatability: ± 0.01% FSO (0.05% 40 K)
Zero Balance: ± 1% FSO
Operating Temp Range: -40 to 93°C (-40 to 200°F)
Compensated Temp Range: 17 to 71°C (60 to 160°F)
Thermal Effects:
Zero: 0.002% FSO/°C
Span: 0.002% FSO/°C
Safe Overload: 150% of capacity
Ultimate Overload: 300% of capacity
Input Resistance: 350 ± 10 ?
Output Resistance: 350 ± 10 ?
Full Scale Deflection: 0.25 to 0.51 mm (0.010 to 0.020")
Construction: 17-4 PH stainless steel
Electrical (LC101/LCM101): 4-Conductor Shielded Cable
<100 kgf/250 lb: 9 m (30') 24 AWG
250 to 1000 kgf/250 to 2000 lb: 9 m (30') 20 AWG
?1500 kgf/3000 lb: 4.5 m (15') 20 AWG
Mating Connector (LC111/LCM111):
?100 kgf/200 lb: PT06F8-4S, sold separately
?250 kgf/250 lb: PT06F10-6S, sold separately
SPECIFICATIONS
Excitation: 10 Vdc, 15 Vdc max
Output: 3 mV/V ± 0.0075 mV/V
Linearity: ± 0.03% FSO (0.1% 40 K)
Hysteresis: ± 0.02% FSO (0.1% 40 K)
Repeatability: ± 0.01% FSO (0.05% 40 K)
Zero Balance: ± 1% FSO
Operating Temp Range: -40 to 93°C (-40 to 200°F)
Compensated Temp Range: 17 to 71°C (60 to 160°F)
Thermal Effects:
Zero: 0.002% FSO/°C
Span: 0.002% FSO/°C
Safe Overload: 150% of capacity
Ultimate Overload: 300% of capacity
Input Resistance: 350 ± 10 ?
Output Resistance: 350 ± 10 ?
Full Scale Deflection: 0.25 to 0.51 mm (0.010 to 0.020")
Construction: 17-4 PH stainless steel
Electrical (LC101/LCM101): 4-Conductor Shielded Cable
<100 kgf/250 lb: 9 m (30') 24 AWG
250 to 1000 kgf/250 to 2000 lb: 9 m (30') 20 AWG
?1500 kgf/3000 lb: 4.5 m (15') 20 AWG
Mating Connector (LC111/LCM111):
?100 kgf/200 lb: PT06F8-4S, sold separately
?250 kgf/250 lb: PT06F10-6S, sold separately
PDFs & Manuals
Show Ratings & Reviews
Rated 4 out of
5
by
Tim H from
good value load cells
I have 3 of these now (different capacities) . the price is reasonable, Omega is a trusted source and the service and support is very good. Make sure to buy the combined tension-compression for the best usefulness. For power and display I use the Digital Panel meter from Automation direct (yes a competitor) but Omega has a similar unit. Buy extra connectors at Newark and the load cells are plug and play (just change the settings on the display). I would give a 5 star but the spot welded covers for the strain gauge are poor in that they can bend and open up. I tie wrap them shut to prevent any issues. The hardware (buttons and rod ends) are useful but a bit pricey.
Date published: 2020-04-19
I'm looking to buy the LCM101-10 and I'm wondering if it is available with a 5m or 10m cable length. Also looking for an electrical schematic, and 3D STEP files. Thank you
Our Tech Support will be in contact and send you an email with a step file. For schematic, see the wiring code below:
Wire Color/Signal
Red/+ INPUT
Black/- INPUT
Green/+ OUTPUT
White/- OUTPUT
Yes, we can do different cable lengths for the LCM111; For the LCM111-10, the standard cable length is 30 ft. ( ~ 10m). You would not need a cable assembly for this load cell. Please call or email us anytime if you have any further questions. Thank you.
Date published: 2019-05-16
Hello, I was curious what thread type is used for attaching any of the accessories or other type of mounting equipment? Thanks.
Thanks for your question?
The threads on both sides of the load cell are M6 x 1.00 female.
hope this was helpful?
Date published: 2019-07-16
Hello I would like to know the excitation voltage of your load cells, I have a datalogger that provides 3.0 V of bridge excitation for powering the bridge. Thank you
Dear Jesus, you must be able to drive the load cells with 3.0 Volts DC. The Load cell sensitivity is 3.0 mV/V nominal, so you will get 9.0 mV at max load at 3.0 Volts DC excitation.
Date published: 2020-02-12