Skip to search Skip to main content Skip to navigation

Welcome to our new website, found an issue or bug? Please report it here

We noticed you are from . Visit your local site for regional offers and live support.

PT100-CY82-CYW4-CY100

Cryogenic Platinum RTD Temperature Sensor

  • Platinum Resistance Sensors for Cryogenic Measurements
  • 4-Wire stainless steel probe
    (Band 4 Silicon Diode Sensor)
  • Special Low Thermal Conductivity
    Extension Wire for CY7 Series Diodes
  • Special Low Thermal Conductivity
    Extension Wire for CY7 Series Diodes
PT100-CY82-CYW4-CY100
From C$ 284.43
Cryogenic Platinum RTD Temperature Sensor PT100-CY82-CYW4-CY100 PT100-CY82-CYW4-CY100
loading all models
Product Overview
  • Electrical Connection Stripped Leads
  • Insulation Tetrafluoroethylene (TFE)
  • Current, Max 0.5 mA
  • Current, Max Note Higher measurement currents can be applied as long as self heating does not change the measurement value more than the needed measurement accuracy. But should not exceed 2mA for 1002 and 0.5mA for 10002.
  • Nominal Resistance 100 Ω
  • Process Temperature, Max 600 °C
  • Recommended Current 0.3 mA
  • Sensor Diameter 1.8 mm
  • Sensor Length 12.1 mm
  • Temperature Sensor Type RTD
Typical connection for Cryogenic sensors such as the CY7 series. The connection points (A) are made close to the sensor in the case of 4-lead measurement, and near or at the instrumentation for a 2-lead measurement. Generally, the 4-lead connection is preferred.
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 3ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;

Can you siggest suitable length extension part for PT-103 by which I can extend up to 2 metres?

Asked by: Sundar
Adding additional wire would make it a custom part so the best approach is to send an rfq to temp@omega.com to get pricing and availability.
Answered by: ApplicationEngineerANB
Date published: 2022-08-25

I would like to ask what is the accuracy of this sensor and also what would be the time constant of the sensor

Asked by: hari
Hi Hari Thermal response time PT-102 & PT-103: 1.75 s at 77 K, 12.5 s at 273 K; Class B accuracy Best regards OMEGA
Answered by: NorwalkAE_RS
Date published: 2020-10-13

Hi We bought recently you thermometer PT-102. Please can you remind me if it is Class A or Class B?

Asked by: Vladislav
Hi Vladislav, Thank you for the inquiry. PT102 is class B accuracy. Best regards OMEGA
Answered by: NorwalkAE_RS
Date published: 2021-12-23
  • y_2025, m_2, d_20, h_9CST