PX2760

High Accuracy, Barometric Pressure Transducers

  • Economical Electric Barometer
  • Rugged Capacitive Design
  • Excellent Long Term Stability
  • Fast Response Time
PX2760
From C$ 959.97
High Accuracy, Barometric Pressure Transducers PX2760 PX2760
loading all models
Product Overview
  • Acceleration, Max 10 g
  • Accuracy Note ± 0.25% span RSS at constant temperature (includes linearity, repeatability and hysteresis)
  • Cable Length 0.6 m
  • Compensated Temperature, Max 55 °C
  • Compensated Temperature, Min 0 °C
  • Electrical Connection Cable
  • Electrical Output 0.1 to 5.1 V
  • Electrical Output Note 0.1 to 5.1 Vdc, 3-wire
  • Materials, Wetted Alumina, Elastomer, Gold, Stainless Steel
  • Media Compatibility Air, Argon, Carbon dioxide, Helium, Hydrogen, Nitrogen, Non-corrosive Dry Gasses, Oxygen
  • Operating Temperature, Max 80 °C
  • Operating Temperature, Min -20 °C
  • Output Resistance <5 Ω
  • Output Signal Voltage
  • Power Supply <0.2 W
  • Pressure Note Range Max Pressure 600 to 1100 mbar 20 psia (450 to 825 mmHg) 800 to 1100 mbar 20 psia (600 to 825 mmHg) 0 to 20 psia 30 psia (0 to 1034 mmHg)
  • Process Connection Size 1/8 in
  • Process Connection Type Barbed Fitting
  • Resolution Infinite, limited only by noise level (0.005% FS)
  • Response Time < 10 ms
  • Response Time Note < 10 ms to reach 90%
  • Sensing Element Capacitive
  • Shock 50 g, operating, 1 ⁄2 sine 10 ms
  • Stability ± 0.25%
  • Stability Note Long-Term Stability: ± 0.25% FS over 6 months @ 70°F
  • Storage Temperature, Max 120 °C
  • Storage Temperature, Min -55 °C
  • Supply Power Note 12 Vdc (9 to 14.5 Vdc max)
  • Supply Voltage 9 to 14.5 Vdc
  • Thermal Effect on Span error, Max ± 1% gage
  • Thermal Effect on Zero error, Max ± 0.01%/°F span , ± 0.01%/°F span
  • Vibration 2 g from 5 to 500 Hz
The PX276 is a rugged duct pressure transmitter that offers up to 3 field-selectable ranges in one unit. The PX276 incorporates a NEMA 4 (IP65) enclosure and a static pressure probe in one unit for ease of installation and accurate static duct pressure readings.

SPECIFICATIONS
Excitation: 12 to 40 Vdc
Output: 4 to 20 mA (2 wire)
Accuracy: ± 1.0% FS
Operating Temperature: -18 to 80°C (0 to 175°F)
Compensated Temperature: -4 to 65°C (25 to 150°F)
Thermal Effects: 0.02%URL/°C (± 0.0125% URL/°F)
Proof Pressure: 10 psi
Media Compatibility: Clean, dry air or inert gas
Termination: Screw terminal block
Enclosure: Enamel-coated, 18 GA steel enclosure with 6061T aluminum probe; NEMA 4 (IP65) rating
Weight: 454 g (1.0 lb)
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 3ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;

Is there an option to have it with digital screen to read out the readings?

Asked by: HossXP
We offer digital displays sold separately that will both power this sensor and display pressure in whatever units of measure you require. A moderately priced display would be our DP8PT. You can locate specifications for this display on our website.
Answered by: JK Applications Engineer
Date published: 2024-07-15
  • y_2025, m_2, d_22, h_8CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.42
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_1
  • loc_en_CA, sid_PX2760-20A5V, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 24ms
  • QUESTIONS, PRODUCT
en-CA
en_CA
Save Configuration