1-800-663-4209 | Your feedback matters, take a quick survey Website Feedback

PHCN-85-Controller
pH/ORP Transmitter and Controller with TFT color display
PHCN-85-Controller
From
C$
1 162.80

Product Overview
- Cutout Dimensions 91 × 91 mm
- Data Interface RS485
- Display Type TFT
- Display Type Note 3.5", 320 × 240 TFT Color Display, Virtual Pointer or Digital Style, Multi-Parameter
- Enclosure Ratings IP65, NEMA Type 4X
- Height 96 mm
- Length 94 mm
- Measurement Resolution pH: 0.01, ORP: mV, 0.1 mV
- Mounting Orientation Panel Mount
- Number of Relays None
- Operating Humidity 0 to 85% RH
- Operating Temperature, Max 50 °C
- Operating Temperature, Min 0 °C
- ORP Accuracy ± 5 mV
- ORP Measurement Range -999 to +999 mV
- pH Accuracy ± 0.1 pH
- PH Measurement Range 0 to 14.00 pH
- Power Consumption 5.5 W
- Process Temperature, Max 0 °C
- Process Temperature, Min 100 °C
- Relay Note Contact photoelectric relay output, programmable; temperature, pH/ORP output can be set for pH/ORP/temperature or timing ON/OFF or PID control Relay: Resistance load 50 mA maximum AC/DC 30V
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -20 °C
- Temperature Accuracy ± 0.5 °C
- Temperature Resolution 0.1 °C
- Temperature Sensor Type 1000 Ω RTD
- Width 96 mm
The new PHCN-85 is the latest edition to the pH/ORP meters from OMEGA. This transmitter and controller offers a large TFT color display and 32-bit processor/ARM core operation system.
The PHCN-85 has two 4 to 20 mA outputs to control or retransmit the pH or ORP and temperature value. In addition, this powerful instrument has 3 relay outputs for pH, ORP, temperature or timing.
The PHCN-85 is also packed with standard features like RS485 communications, a calendar function and password protection. The power supply input even has automatic polarity detection.
Specifications
Measurement:
pH:
Range: 0.00 to 14.00
Resolution: 0.01
Accuracy: ±0.1
ORP:
Range: -1999 to +1999
Resolution: mV, 0.1mV
Accuracy: ±5 mV (display)
Temperature:
Range: 0 to 100 °C (32 to 212 °F)
Resolution: 0.1
Accuracy: ±0.5
Automatic Temperature Compensation: 0 to 50 °C (32 to 122 °F)
Temperature Input: 1000 Ω RTD
Current Outputs (2): Isolated, adjustable, reversible, 4 to 20 mA output, control or transmitter mode
Loop Resistance: 400 Ω maximum, DC 24V
Current Output: 4 to 20 mA
Resolution: ±0.1 mA
Relay Outputs (3): Contact photoelectric relay output, programmable; temperature, pH/ORP output can be set for pH/ORP/temperature or timing ON/OFF or PID control
Relay: Resistance load 50 mA maximum AC/DC 30V
Data Communication: RS485 Modbus protocol
Power Supply: DC 24V ±4V
Consumption: <5.5 W
Working Environment
Temperature: 0 to 50 °C (32 to 122 °F)
Relative Humidity: ≤85% RH (non-condensing)
Storage Temperature: -20 to 60 °C (-4 to 140 °F)
Relative Humidity: ≤85% RH (non-condensing)
Protection Level: NEMA 4 (IP65) with rear cover
Dimensions: 96 H × 96 W × 94 D mm (3.78 x 3.78 x 3.7")
Panel Cutout: 91 × 91 mm (3.58 x 3.58")
The PHCN-85 has two 4 to 20 mA outputs to control or retransmit the pH or ORP and temperature value. In addition, this powerful instrument has 3 relay outputs for pH, ORP, temperature or timing.
The PHCN-85 is also packed with standard features like RS485 communications, a calendar function and password protection. The power supply input even has automatic polarity detection.
Specifications
Measurement:
pH:
Range: 0.00 to 14.00
Resolution: 0.01
Accuracy: ±0.1
ORP:
Range: -1999 to +1999
Resolution: mV, 0.1mV
Accuracy: ±5 mV (display)
Temperature:
Range: 0 to 100 °C (32 to 212 °F)
Resolution: 0.1
Accuracy: ±0.5
Automatic Temperature Compensation: 0 to 50 °C (32 to 122 °F)
Temperature Input: 1000 Ω RTD
Current Outputs (2): Isolated, adjustable, reversible, 4 to 20 mA output, control or transmitter mode
Loop Resistance: 400 Ω maximum, DC 24V
Current Output: 4 to 20 mA
Resolution: ±0.1 mA
Relay Outputs (3): Contact photoelectric relay output, programmable; temperature, pH/ORP output can be set for pH/ORP/temperature or timing ON/OFF or PID control
Relay: Resistance load 50 mA maximum AC/DC 30V
Data Communication: RS485 Modbus protocol
Power Supply: DC 24V ±4V
Consumption: <5.5 W
Working Environment
Temperature: 0 to 50 °C (32 to 122 °F)
Relative Humidity: ≤85% RH (non-condensing)
Storage Temperature: -20 to 60 °C (-4 to 140 °F)
Relative Humidity: ≤85% RH (non-condensing)
Protection Level: NEMA 4 (IP65) with rear cover
Dimensions: 96 H × 96 W × 94 D mm (3.78 x 3.78 x 3.7")
Panel Cutout: 91 × 91 mm (3.58 x 3.58")
PDFs & Manuals