OM-CP-RTDTEMP2000-Logger
Precision RTD Temperature Data Logger with Display
Models In Stock
- Large, Backlit LCD Display
- Front Keypad with Lock Feature
- High Speed Download
- Real-Time Operation
OM-CP-RTDTEMP2000-Logger
From
C$
1 001.68
Models In Stock

Product Overview
- Battery life up to 1 yr
- Battery Size and Type Li-ion, 9V
- Calibration Digital calibration through software
- Data Interface Note PC serial or USB (interface cable required);115,200 baud
- Display Type LCD
- Display Type Note DOT-Matrix LCD , 63 x 35 mm
- Electrical Connection Screw Terminal
- Enclosure Aluminium
- Height 32 mm
- Length 122 mm
- Measurement Type Temperature
- Memory 174762
- Memory Note 174,762 total readings
- Number of Measurements 1
- Operating Temperature, Max 60 °C
- Operating Temperature, Min -20 °C
- Probe Calibrated Accuracy ± 0.05°C
- Process Temperature, Max 850 °C
- Process Temperature, Min -200 °C
- Relative Humidity, Max 0.95
- Relative Humidity, Min 0%
- Samples,Max 175000
- Sampling Rate 0.5 Hz (2 seconds between updates)
- Sampling Rate Note 1 reading every 2 seconds to 1 every 24 hours
- Start Modes Adjustable Delay Start, Immediate Start
- Supply Power Type Battery
- Temperature Resolution 0.01 °C
- Time Accuracy ± 1 minute/month
- Waterbox Depth 39 mm
- Waterbox Height 74 mm
- Waterbox Width 148 mm
- Weight 440 g
- Width 84 mm
The OM-CP-RTDTEMP2000 is a battery powered precision RTD temperature recorder, with a LCD display. It is perfect for applications requiring precise temperature readings. The 8-button keypad and large LCD provide convenient access to current data and recorder setup. Additionally, memory and battery levels, external power status, and sampling and recording status are shown on the LCD.
Available on-screen data includes: statistics (min, max, average); recording status (start, stop and recording rate); display options (units, text size); and calibration information (date calibrated, date for recalibration). Statistics can be cleared at any time during logging. With 174,762 readings, the OM-CP-RTDTEMP2000 has one of the largest memory capacities of any similar data recorder on the market. The non-volatile memory will retain recorded data, even if AC and battery power are lost.
The OM-CP-RTDTEMP2000 runs on a 9 V battery with a typical battery life of 1 year at a 1 min reading rate with the display off. With continuous LCD and no backlight usage, the average battery life is 30 days. For power savings, both the LCD and backlight have configurable auto-off options. Those wanting to keep the LCD display and backlight continuously active need only to connect the unit to an AC power supply, keeping the 9 V battery connected as a back-up. NIST traceable calibration is available for users needing to meet regulatory requirements. Creating permanent records, performing data calculations, and the graphing of data is quick and easy: Simply connect the interface cable to an available USB or serial port and, with a few clicks of the mouse, data is downloaded and ready for review or export to Excel.
Specifications
TEMPERATURE PROBE
Measurement Range: -200 to 850°C (-328 to 1562°F)
Input Type: 100 Ω Pt RTD, α = 0.00385
Input Connection: 2, 3, or 4-wire scew terminal block
Resolution: 0.01°C
Probe Calibrated Accuracy: ± 0.05°C (-200 to 260°C) ± 0.3°C (260 to 850°C)
Specified Accuracy Range: -200 to 850°C (-328 to 1562°F)
DOT-MATRIX LCD
Dimensions: 63 mm x 35 mm (2.5 x 1.375')
Text: Configurable channel text size
Indicators: Power, status, memory
Backlight: Configurable w/auto shut-off and contrast adjustment
Start/Stop Time: Software programmable start time
and date, up to six months in advance; programmable stop time
Memory: 174,762 total readings; software configurable memory wrap
Reading Rate: 1 reading every 2 seconds to 1 every 24 hours
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Battery Type: 9 V lithium battery included, user replaceable; optional AC adapter
Battery Life: 1 year battery life at 1 min reading rate with display
off. 30 days typical with continuous LCD and no backlight usage.
Data Format: Date and time stamped °C, °F, K, °R
Time Accuracy: ± 1 minute/month (at 20° to 30°C)
Computer Interface: PC serial or USB (interface cable required); 115,200 baud
Software: XP SP3/Vista/7 and 8 (32-bit and 64-bit)
Operating Environment: -20 to 60°C, (-4 to 140°F) 0 to 95% RH non-condensing
Dimensions: 122 x 84 x 32 mm (4.8 x 3.3 x 1.25')
Weight: 440 g (16 oz)
Enclosure: Black anodized aluminum
Available on-screen data includes: statistics (min, max, average); recording status (start, stop and recording rate); display options (units, text size); and calibration information (date calibrated, date for recalibration). Statistics can be cleared at any time during logging. With 174,762 readings, the OM-CP-RTDTEMP2000 has one of the largest memory capacities of any similar data recorder on the market. The non-volatile memory will retain recorded data, even if AC and battery power are lost.
The OM-CP-RTDTEMP2000 runs on a 9 V battery with a typical battery life of 1 year at a 1 min reading rate with the display off. With continuous LCD and no backlight usage, the average battery life is 30 days. For power savings, both the LCD and backlight have configurable auto-off options. Those wanting to keep the LCD display and backlight continuously active need only to connect the unit to an AC power supply, keeping the 9 V battery connected as a back-up. NIST traceable calibration is available for users needing to meet regulatory requirements. Creating permanent records, performing data calculations, and the graphing of data is quick and easy: Simply connect the interface cable to an available USB or serial port and, with a few clicks of the mouse, data is downloaded and ready for review or export to Excel.
Specifications
TEMPERATURE PROBE
Measurement Range: -200 to 850°C (-328 to 1562°F)
Input Type: 100 Ω Pt RTD, α = 0.00385
Input Connection: 2, 3, or 4-wire scew terminal block
Resolution: 0.01°C
Probe Calibrated Accuracy: ± 0.05°C (-200 to 260°C) ± 0.3°C (260 to 850°C)
Specified Accuracy Range: -200 to 850°C (-328 to 1562°F)
DOT-MATRIX LCD
Dimensions: 63 mm x 35 mm (2.5 x 1.375')
Text: Configurable channel text size
Indicators: Power, status, memory
Backlight: Configurable w/auto shut-off and contrast adjustment
Start/Stop Time: Software programmable start time
and date, up to six months in advance; programmable stop time
Memory: 174,762 total readings; software configurable memory wrap
Reading Rate: 1 reading every 2 seconds to 1 every 24 hours
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Battery Type: 9 V lithium battery included, user replaceable; optional AC adapter
Battery Life: 1 year battery life at 1 min reading rate with display
off. 30 days typical with continuous LCD and no backlight usage.
Data Format: Date and time stamped °C, °F, K, °R
Time Accuracy: ± 1 minute/month (at 20° to 30°C)
Computer Interface: PC serial or USB (interface cable required); 115,200 baud
Software: XP SP3/Vista/7 and 8 (32-bit and 64-bit)
Operating Environment: -20 to 60°C, (-4 to 140°F) 0 to 95% RH non-condensing
Dimensions: 122 x 84 x 32 mm (4.8 x 3.3 x 1.25')
Weight: 440 g (16 oz)
Enclosure: Black anodized aluminum
PDFs & Manuals
Software & Drivers
Show Ratings & Reviews
Can the RS232 port be used to communicate with the unit to read the current temperature, etc? Or is it only used with the bundled software? The manual seems to suggest the latter. The former would make this immensely more useful.
Thank you for your question. Just with the supplied software.
Date published: 2023-02-23
There seems to be a disclaimer that the logger may not work properly with 3 wire RTDs in the manual, is this right? Can this be used with 3 wire? I'm only confused because as far as I know, 3W is better than 3W.
Thank you for your question. 3-wire Rtd will work fine with the logger.
Date published: 2021-11-08