Skip to search Skip to main content Skip to navigation

Bienvenue sur notre nouveau site Web, vous avez rencontré un problème ou un bug? Veuillez le signaler ici

Nous avons remarqué que vous veniez de . Visitez votre site local pour connaître les offres régionales et l'assistance en direct.

HH25U-Series

1 Channel High Accuracy 0.1% Type K, J, T Thermocouple Meter

  • Auto Power-Off with Disable Feature (15 Minutes)
  • Single Thermocouple Input
  • Large LCD Screen
  • Data Hold Function Freeze Readings on Display.
1 Channel High Accuracy 0.1% Type K, J, T Thermocouple Meter HH25U-Series HH25U-Series
loading all models
L'aperçu du produit
  • Additional Features 1 MIN/MAX, Auto Power off
  • Electrical Connection Miniature Male Thermocouple Connector
  • Functions Max/Min. hold
  • Height 159 mm
  • Length 33 mm
  • Number of Channels 1
  • Sampling Rate 3 Hz (3 updates/second)
  • Supply Power Note 9V (included), MN 1604
  • Supply Voltage 9 Vdc
  • Temperature Coefficient 0.1 times the applicable accuracy specification per °C from 0 to 18°C and 28 to 50°C (32 to 64°F and 82 to 122°F)
  • Temperature Sensor Type Thermocouple
  • Thermocouple Type J, K, T
  • Weight 170 g
  • Weight Note 170 g (6 oz)
  • Width 55 mm

Description

The HH-25U handheld thermocouple thermometers feature Type K, J, T thermocouple inputs and are available in single input models. The HH-25U is a single-input model that has an auto range resolution of 0.1 or 1°. are °C/°F switchable, and maximum/Minimum reading/display hold functions and low battery indicator. Offered at an economical price.

Specifications
Input: Single K/J/T
Temperature Range K: -200 to 1370ºC (-328 to 2498ºF)
Resolution (K): 0.1ºC (0.1ºF), over 1000ºC 1ºC/1ºF
Accuracy (K): ± (0.1% rdg + 0.5°C) for -50 to 1370°C others ± (0.1% rdg + 2°C)
Temperature Range J: -80 to 760ºC (-112 to 1400ºF)
Accuracy (J): ± 1% of rdg + 1ºC/2ºF
Temperature Range T: -80 to 400ºC (-112 to 752ºF)
Accuracy (T): ± 1% of rdg + 1% 200 to 400ºC, others ± 0.4% rdg + 0.6 ºC/1.2ºF
Temperature Coefficient: 0.1 times the applicable accuracy specification per °C from 0 to 18°C and 28 to 50°C (32 to 64°F and 82 to 122°F)
Function: Max/Min. hold
Power: 9V battery
Dimensions: 159 L x 55 W x 38 D mm (6.25 x 2.16 x 1.49")
Weight: 170 g (6 oz)
Operating Temperature: -10 to 50ºC (14 to 122ºF)
Rated 5 out of 5 by from Highly recommend omega products Great product very reliable instrument that we use daily.
Date published: 2023-07-26
Rated 5 out of 5 by from 5 star Works as advertised. The shipping was very fast. Good experience.
Date published: 2023-04-10
Rated 5 out of 5 by from Perfect for my requirements The thermometer worked perfectly with all my T type thermocouples and provided me with fast temperature readouts. Once the unit is set for my T type thermocouples I never have to reset it.
Date published: 2021-02-25
  • y_2025, m_2, d_22, h_8
  • bvseo_bulk, prod_bvrr, vn_bulk_3.0.42
  • cp_1, bvpage1
  • co_hasreviews, tv_0, tr_3
  • loc_en_CA, sid_HH-25U, prod, sort_[SortEntry(order=HELPFULNESS, direction=DESCENDING), SortEntry(order=POSITIVE_FEEDBACK, direction=DESCENDING), SortEntry(order=SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 5ms
  • REVIEWS, PRODUCT

My unit is stuck at 1400 reading. Is this a common issue that can be fixed?

Asked by: Darsl
Thank you for your inquiry. Keep in mind, that Temperature Range J is -80 to 760ºC (-112 to 1400ºF)
Answered by: Engineer CB
Date published: 2021-04-14

What type of input connector does the unit has; standard or mini?

Asked by: Loren
Thank you for your inquiry. It will accept mini male connector.
Answered by: Application Engineer SP
Date published: 2023-06-21

What is the measurement rate?

Asked by: Gior
The sampling rate is 0.75 seconds for the HH-25U meter.
Answered by: ApplicationEngineerANB
Date published: 2022-04-12

Does this unit come with a calibration certificate and if so, does the date start when first put into use?

Asked by: SteveR
The meter does come with a C of C standard and there isn't any expiration date for it. Omega recommends annual calibration on meters and more often if they are used in critical applications.
Answered by: ApplicationEngineerANB
Date published: 2021-04-09

Hello, Is this device compatible with thermocouple slides? We monitor the temperature on our Thermobrites daily for FISH analysis.

Asked by: Armesegedden
The HH-25U will accept Thermocouple mv signals and interpret it into temperature readings from a type K,J, or T. If the Thermobrites have Thermocouple outputs and miniature connectors then it would accept and display temperature for them.
Answered by: ApplicationEngineerANB
Date published: 2023-10-26

What kind of thermocouple does it come with?

Asked by: james1989
The HH25U does not come with a Thermocouple so you can search the 5SC or 5SRTC series for wire probes with the appropriate connector.
Answered by: ANBENG
Date published: 2021-04-14

where are the other specs on this unit? Accuracy per TC type, measurement rate, etc.

Asked by: engtech
Thank you for your question. Specifications for this unit are below: Input: Single K/J/T Temperature Range K: -200 to 1370ºC (-328 to 2498ºF) Resolution (K): 0.1ºC (0.1ºF), over 1000ºC 1ºC/1ºF Accuracy (K): ±(0.1% rdg + 0.5°C) for -50 to 1370°C; others ±(0.1% rdg + 2°C) Temperature Range J: -80 to 760ºC (-112 to 1400ºF) Accuracy (J): ±1% of rdg + 1ºC/2ºF Temperature Range T: -80 to 400ºC (-112 to 752ºF) Accuracy (T): ±1% of rdg + 1% 200 to 400ºC, others ±0.4% rdg + 0.6 ºC/1.2ºF Temperature Coefficient: 0.1 times the applicable accuracy specification per °C from 0 to 18°C and 28 to 50°C (32 to 64°F and 82 to 122°F) Function: Max/Min. hold Power: 9V battery Dimensions: 159 L x 55 W x 38 D mm (6.25 x 2.16 x 1.49") Weight: 170 g (6 oz) Operating Temperature: -10 to 50ºC (14 to 122ºF)
Answered by: Application Engineer
Date published: 2022-04-04
  • y_2025, m_2, d_21, h_8CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.42
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_7
  • loc_en_CA, sid_HH-25U, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 34ms
  • QUESTIONS, PRODUCT
fr-CA
fr
Enregistrer la configuration