Bienvenue sur notre nouveau site Web, vous avez rencontré un problème ou un bug? Veuillez le signaler ici

LC901
Bolt Sensors with Mounting Washers
Modèles en stock
- All Stainless Steel Construction
- Mounting Washers Included
- NIST Calibration Certificate
LC901
Depuis
C$
686.00
Modèles en stock

L'aperçu du produit
- Accuracy ± 3.5 %
- Bridge Resistance 120 Ω
- Calibration Option 2 POINT NIST
- Electrical Output 2 mV/V
- Force Type Compression
- Hysteresis ± 3.5%
- Linearity ± 3.5%
- Load Cell Type Load Washer
- Materials Stainless Steel
- Operating Temperature, Max 93 °C
- Operating Temperature, Min -54 °C
- Overload, Max 150% FS
- Process Temperature, Max 71 °C
- Process Temperature, Min 16 °C
- Repeatability ± 3.5%
- Safe Overload 150% FS
- Supply Power Note 10 Vdc, 15 Vdc maximum
- Supply Voltage 10 Vdc
- Thermal Span Effect ± 0.036% rdg/°C
- Thermal Zero Effect ± 0.018% FSO/°C
- Zero/Balance Error ± 4%
OMEGA’s LC900 Series bolt force sensor is a compression load cell designed to measure the clamping force of a bolt. With excellent long term stability and all stainless steel construction, the LC900 Series delivers high reliability in severe industrial environments. This unit is designed to operate by mounting on a flat surface between two mounting washers (included).
The LC901 is also available in metric configurations, with ranges from 0-10,000 to 0-500,000 N. Please see the LCM901 for complete details, or contact our Pressure department for more information.
SPECIFICATIONS
Excitation: 5 Vdc (10V max)
Output: 2 mV/V nominal
Accuracy: ± 3.5% FSO (linearity, hysteresis and repeatability combined)
Zero Balance: ± 4% FSO
Operating Temp Range: -54 to 93°C (-65 to 200°F)
Compensated Temp Range: 16 to 71°C (60 to 160°F)
Thermal Effects:
Zero: ± 0.018% FSO/°C
Span: ± 0.036% rdg/°C
Safe Overload: 130% of capacity
Ultimate Overload: 150% of capacity
Bridge Resistance: 120 ? nominal
Construction: Stainless steel
Electrical Connection: 1.2 m (4') insulated shielded cable
The LC901 is also available in metric configurations, with ranges from 0-10,000 to 0-500,000 N. Please see the LCM901 for complete details, or contact our Pressure department for more information.
SPECIFICATIONS
Excitation: 5 Vdc (10V max)
Output: 2 mV/V nominal
Accuracy: ± 3.5% FSO (linearity, hysteresis and repeatability combined)
Zero Balance: ± 4% FSO
Operating Temp Range: -54 to 93°C (-65 to 200°F)
Compensated Temp Range: 16 to 71°C (60 to 160°F)
Thermal Effects:
Zero: ± 0.018% FSO/°C
Span: ± 0.036% rdg/°C
Safe Overload: 130% of capacity
Ultimate Overload: 150% of capacity
Bridge Resistance: 120 ? nominal
Construction: Stainless steel
Electrical Connection: 1.2 m (4') insulated shielded cable
PDF et manuels
Afficher les notes et les avis
Rated 3 out of
5
by
Anonymous from
120 Ohm Load Cells
The load cells have 120 Ohm resistance. Many of the load cell amplifiers available, including the one recommended for this product, require the load cell to have at least 300 Ohms resistance. That makes this device hard to use.
Date published: 2020-10-26
What is the actual transducer height without the additional washers?
MLFEP, Thank you for your question. Please open the spec sheet and review the "W" dimension corresponding to your range.
Date published: 2020-07-01
What is the outside diameter of the signal cable and what is the wire gauge of its conductors?
dmeigs LC901 wire gauge is 24 AWG, for further questions kindly email pressure@omega.com
Thank you
Date published: 2020-07-03
Are the top and bottom washers required if the bottom of the transducer sits on a flat surface, and the item on the top has a .750 O.D. and .266 I.D.?
The top washer should always be used. It assures the load is distributed around the I.D. evenly. The bottom washer would not be need if the surface the surface the load is resting on is perfectly flat
Date published: 2021-11-08