Bienvenue sur notre nouveau site Web, vous avez rencontré un problème ou un bug? Veuillez le signaler ici

CDTX-111-CDTX-112-Series
Panel Mount Conductivity and Temperature Monitor and Controller
Modèles en stock
- On/Off Relay Outputs
- Economical
- Automatic or Manual Temperature Compensation
- Selectable Zero and Span Analog Output
CDTX-111-CDTX-112-Series
Depuis
C$
377.49
Modèles en stock

L'aperçu du produit
Industrial process conductivity/temperature controllers CDTX-111 and CDTX-112 are panel instruments for in-process monitoring of conductivity. Range switch over and cell constant checks can both be freely set and adjusted during operation. Unique signal collection and processing technology, and a special thick film circuit are employed to ensure accurate measurements and stable operation. Other features include linearized data, automatic temperature compensation, and maintenance-free cells. The control functions trigger relays based on conductivity setpoints. The CDTX-111 and CDTX-112 are ideal auxiliary instruments for many types of water testing applications.
Specifications
Display Type
CDTX-111: LED, 13 mm height
CDTX-112: LCD, 13 mm height
Measurement: 0 to 18 M? or 0 to 19.99 µS/cm
Range (Depending upon cell constant)
CDTX-111: 0 to 199.9 µS/cm; 0 to 1999 µS/cm
CDTX-112: 0 to 999.9 µS/cm; 0 to 9999 µS/cm, 0 to 100mS/cm, 0 to 600 mS/cm
Temperature: 0 to 100°C (32 to 212°F)
Resolution: 0.01 M?
CDTX-111: 0.01 to 1 µS/cm
CDTX-112: 0.01 to 0.01 mS/cm
Temperature: 0.1°C (32.1°F)
Accuracy: ±1% FS, ± 0.2°C
Temperature: Automatic/manual
Compensation: 0 to 100°C (32 to 212°F)
Output Relays: 2 programmable, 10 A
Control Type: On/off
Relay Set Point Hysteresis: User programmable
Current Output
Type: Isolated 4 to 20 mA
Range: Selectable zero and span
Accuracy: ±0.02 mA
Maximum Load: 750 ?
Power Supply: AC 110V to 220V ±10%, 50/60 Hz
Weight
CDTX-111: 580 g (20.4 oz)
CDTX-112: 650 g (22.9 oz)
Dimensions
CDTX-111: 96 × 48 × 110 mm (3.78 x 1.89 x 4.33")
CDTX-112: 96 × 96 × 110 mm (3.78 x 3.78 x 4.33")
Mount Type: Panel
Cut-Out Size
CDTX-111: 92 × 44 mm (3.6 x 1.73")
CDTX-112: 92 × 92 mm (3.6 x 3.6")
Enclosure Rating: IP54
Environmental Operating Ambient Temperature: -10 to 55°C (14 to 131°F)
Requirements
Relative Humidity: 5 to 95% non-condensing
Maximum Operating Altitude: 3000 m (10,000')
Specifications
Display Type
CDTX-111: LED, 13 mm height
CDTX-112: LCD, 13 mm height
Measurement: 0 to 18 M? or 0 to 19.99 µS/cm
Range (Depending upon cell constant)
CDTX-111: 0 to 199.9 µS/cm; 0 to 1999 µS/cm
CDTX-112: 0 to 999.9 µS/cm; 0 to 9999 µS/cm, 0 to 100mS/cm, 0 to 600 mS/cm
Temperature: 0 to 100°C (32 to 212°F)
Resolution: 0.01 M?
CDTX-111: 0.01 to 1 µS/cm
CDTX-112: 0.01 to 0.01 mS/cm
Temperature: 0.1°C (32.1°F)
Accuracy: ±1% FS, ± 0.2°C
Temperature: Automatic/manual
Compensation: 0 to 100°C (32 to 212°F)
Output Relays: 2 programmable, 10 A
Control Type: On/off
Relay Set Point Hysteresis: User programmable
Current Output
Type: Isolated 4 to 20 mA
Range: Selectable zero and span
Accuracy: ±0.02 mA
Maximum Load: 750 ?
Power Supply: AC 110V to 220V ±10%, 50/60 Hz
Weight
CDTX-111: 580 g (20.4 oz)
CDTX-112: 650 g (22.9 oz)
Dimensions
CDTX-111: 96 × 48 × 110 mm (3.78 x 1.89 x 4.33")
CDTX-112: 96 × 96 × 110 mm (3.78 x 3.78 x 4.33")
Mount Type: Panel
Cut-Out Size
CDTX-111: 92 × 44 mm (3.6 x 1.73")
CDTX-112: 92 × 92 mm (3.6 x 3.6")
Enclosure Rating: IP54
Environmental Operating Ambient Temperature: -10 to 55°C (14 to 131°F)
Requirements
Relative Humidity: 5 to 95% non-condensing
Maximum Operating Altitude: 3000 m (10,000')
PDF et manuels
Afficher les notes et les avis
Is it possible to buy the the CDTX-111 monitor with normally closed AL and AH contacts?
Thank you for your inquiry about the conductivity controller. The CDTX-111 unit has (2) 10A relays to control your conductivity range.
Date published: 2021-05-27
About your CDTX-111 conductivity monitor, we need it to monitor the water resistivity in real tome. I have a quetion to ask: Terminal Pin 12 & Pin 13 behind it is connected to the temperature sensor, is this sensor necessary in our application?
Hi, thanks for connecting with Omega. The CDTX-111 conductivity controller has Automatic Temperature Compensation (ATC) and you do not necessarily need it unless the sample you are measuring is changing or fluctuating in temperature. If the liquid measured is at a constant temperature, then, measuring conductivity will not be affected. Please email flow@omega.com with any further questions. Thank you
Date published: 2021-03-16
The manual calls for 1 calibration level, however is it possible to calibrate multiple levels using different known standard solutions? I am having an issue with full range 0-2000us accuracy.
Thank you for your inquiry. You may calibrate using multiple solutions, it will only retain the 1 source and best to check the correct probe cell constant being used for higher conductivity solution.
Date published: 2021-03-16