Bienvenue sur notre nouveau site Web, vous avez rencontré un problème ou un bug? Veuillez le signaler ici

LCWD
Load Washers, Measure Bolt Compression Forces
Modèles en stock
- Ranges to 200,000 lb
- Compact Design
- Spherical Washers Included
- Load Buttons Included
LCWD
Depuis
C$
867.23
Modèles en stock

L'aperçu du produit
- Cable Length 5 ft
- Calibration Option 4 POINT NIST
- Electrical Connection Cable
- Force Type Compression
- Full Scale Deflection 0.001 in
- Load Cell Type Load Washer
- Materials Stainless Steel
- Operating Temperature, Max 127 °C
- Operating Temperature, Min -40 °C
- Overload, Max 200% FS
- Process Temperature, Max 71 °C
- Process Temperature, Min 16 °C
- Safe Overload 120% FS
- Supply Power Note 10 Vdc, 15 Vdc maximum
- Supply Voltage 10 Vdc
- Thermal Span Effect ± 0.018% FSO/°C
- Thermal Zero Effect ± 0.018% FSO/°C
- Wiring/Connector Table Red: (+) Excitation; Black: (-) Excitation; Green: (+) Output; White: (-) Output
- Zero/Balance Error ± 4%
The LCWD Series are for measuring bolt compressional forces in industrial, scientific and development applications. Spherical washers are included to assure proper alignment of the LCWD.
The LCWD is also available in metric configurations, with ranges from 0-100 to 0-200,000 N. Please see the LCMWD for complete details, or contact our Pressure department for more information.
SPECIFICATIONS
Excitation: 10 Vdc (15V max)
Output: 1 mv/V
Accuracy (Linearity and Repeatability Combined):
?10K lb/50 kN: ± 0.5% BFSL
>10K lb/50 kN: ± 0.75% BFSL
Zero Balance: ± 4.0% FSO
Operating Temp Range: -40 to 127°C (-40 to 260°F)
Compensated Temp Range: 16 to 71°C (60 to 160°F)
Thermal Effects:
Zero: ± 0.018% FSO/°C
Span: ± 0.018% FSO/°C
Safe Overload: 120% of capacity
Ultimate Overload: 200% of capacity
Bridge Resistance:
<50k lb/200 kN: 350 ? min
?50k lb/200 kN: 700 ?
Full Scale Deflection: 0.025 mm (0.001") typical
Construction: Stainless steel
Electrical: 1.5 m (5') insulated shielded cable
The LCWD is also available in metric configurations, with ranges from 0-100 to 0-200,000 N. Please see the LCMWD for complete details, or contact our Pressure department for more information.
SPECIFICATIONS
Excitation: 10 Vdc (15V max)
Output: 1 mv/V
Accuracy (Linearity and Repeatability Combined):
?10K lb/50 kN: ± 0.5% BFSL
>10K lb/50 kN: ± 0.75% BFSL
Zero Balance: ± 4.0% FSO
Operating Temp Range: -40 to 127°C (-40 to 260°F)
Compensated Temp Range: 16 to 71°C (60 to 160°F)
Thermal Effects:
Zero: ± 0.018% FSO/°C
Span: ± 0.018% FSO/°C
Safe Overload: 120% of capacity
Ultimate Overload: 200% of capacity
Bridge Resistance:
<50k lb/200 kN: 350 ? min
?50k lb/200 kN: 700 ?
Full Scale Deflection: 0.025 mm (0.001") typical
Construction: Stainless steel
Electrical: 1.5 m (5') insulated shielded cable
Afficher les notes et les avis
Rated 5 out of
5
by
Mike V from
Quick Response
Needed this item quickly for product testing. Was able to have it overnighted and worked perfectly.
Date published: 2023-02-17
What do i need record data from this load cell Item# LCWD-100K
You could use either a DP8EPT Display to visually read the force load and communicate with a computer via USB to collect the data or you can use our OM-CP-BRIDGE101A data logger (no display) to collect data via USB to a computer.
Date published: 2021-09-09
Are the LCWD load cells certified or can they be purchased with certification?
The LCWD load cells are shipped with a NIST traceable calibration report included
Date published: 2024-06-10
When I setup theHigh Accuracy, Load Washers to Measure Bolt Compression with the DP8EPT display, I cannot Tare/ Zero with the display?
Thank you for your inquiry, the tare function must be enabled through the menu on the DP8EPT please consult the manual for instructions.
Date published: 2022-09-21
How do I prevent the load cell from rotating when torqueing a bolt against the cell?
Bobby, once the load cells are mounted in place, it is suggested that you use hardened washers both on the top and bottom of the load cell while torqueing your bolt.
Date published: 2023-09-07
Is the STRAIN GAGE/BRIDGE TRANSDUCER-SIGNAL CONDITIONERS required for a purchase of the High Accuracy, Load Washers to Measure Bolt Compression + DP8EPT display?
Thank you for your inquire
No need to go with a signal conditioner unless you are in need to change the signal for the distance?
most load cell can go as far as 50 feet.
Any distance longer you would need to change the loadcell signal to a current or voltage.
Date published: 2022-07-11
Do I need to buy the custom meter, or can I read this device with an existing standard voltmeter or ADC?
You can use a precise digital to measure the 0 to 20mVDC output signal and then calculate the applied load is. We suggest using one of our DP8EPT displays. It will supply the required 10VDC power to the cell and you will be able scale the display to read directly in units of force.
Date published: 2022-05-17
On the LCWD Load cell, can I use a 5 VDC excitation instead of 10 VDC?
Yes you can. The only difference would be the output signal would be 50% lower
Date published: 2023-11-14