- High measurement temperatures up to 1800 °C (3272 °F)
- Many fields of view options, 2:1,15:1, 20:1, 40:1 and 80:1
- Features a built-in LCD display with keypad
- User selectable options for analog outputs and USB 2.0 digital
OSAO-Series
Depuis
C$
1,333.12
Modèles en stock

L'aperçu du produit
- Cable Length 3 m
- Control Output Relay Output with hysteresis 60V DC/42V AC RMS,0.4A
- Data Interface USB
- Display Yes
- Emissivity, Min 0.1
- Enclosure Ratings IP65
- Housing Height 33 mm
- Housing Length 112.5 mm
- Housing Width 82.5 mm
- Intrisically Safe No
- Isolation Power supply, *Digital output and Analog output are galvanically isolated * Not applicable for USB 2.0 digital output
- Operating Temperature, Min 0 °C
- Power Consumption 2.5 W
- Relative Humidity, Max 0.95
- Relay Output 60 Vdc
- Spectral Response, Min 8 μm
- Storage Temperature, Max 70 °C
- Storage Temperature, Min -20 °C
- Supply Voltage 12 to 28 Vdc
- Weight 600 g
The OSAO-Series high temperature IR sensors are designed to offer peak performance over a wide range of temperatures. The integrated LCD display offers a convenient way to adjust the settings and measure the temperature. The display is capable of setting various parameters like emissivity, analog sub range, set point, hysteresis, analog output, temperature unit, response time, peak picker, sensor address, and more!
The integrated laser targeting light (AO250 Models) and optional mounting accessory significantly speed up installation. Accessories like the Water Cooling Jacket (with air purge) and Air Purge Unit-Sensor Head allow the sensor to be installed in varying environments. If the ambient temperature is over 70 °C (158 °F) or lots of dust can build up on the sensor head then these accessories become more of a necessity. If ambient temperatures are over 120 °C (248 °F) then the AO50H Model with an ambient temperature rating up to 180 °C (356 °F) is a perfect solution.
The OSAO-Series sensors have adjustable emissivity and can measure many different surfaces such as Iron and Steel, Lead, Marble, Brick, Asphalt, and many more!
PDF et manuels
Logiciels et pilotes
Étalonnage
Afficher les notes et les avis
Rated 5 out of
5
by
Shweta Jaroli from
Great Performance
We are satisfied with performance of this pyrometer.
Date published: 2021-09-02
Rated 1 out of
5
by
SSmith from
Does Not Work
The display frequently freezes at a temperature and stops live updating. The power wires going into the device are exposed because the wrapping is too short. They will likely break from moving the device or rip out. The OmegaSoft data collection software will either record to an old file or say that it can't record because a file is active. The "Start" and "Stop" data collection functions do not work.
Date published: 2021-03-26
What length does the cable come in?
Thank you for your inquiry. The cable length of the sensor head is 10ft.
Date published: 2022-10-24
Hi, we have one of these units. Lately it stopped working and shows error code 14. We think the sensor has failed (it was working in a sometimes corrosive environment). Can you buy replacement sensors for these? (0-800 C, 15:1 field of view, 8-14 um)
Hi Mike,
Thank you for the inquiry. Sorry to hear of the difficulty. Please forward this request to TEMP@OMEGA.COM
Best regards
OMEGA
Date published: 2023-05-12
It is possible to order a product with a cable 8 meters long (26,2ft)?
It would be a special request and if you send an email to temp@dwyeromega.com it can be reviewed and priced out if available.
Date published: 2024-02-21
We would like to use this to monitor material surface temperature while heating with an Oxy/Act torch. Will the heat from the torch flame influence the temperature readings on the material surface?
Thank you for your inquiry. We believe the best assistance would come from speaking to us directly at 888-826-6342 ask for the Temperature Department Mon. – Fri. (HOURS 8:30AM - 5:30PM EST).