Bienvenue sur notre nouveau site Web, vous avez rencontré un problème ou un bug? Veuillez le signaler ici

M12TXSS
Stainless Steel RTD probes with 4 to 20 mA analog output & Mounting Threads
- Programmable Transmitter
- M12 Connection
- 4 to 20 mA Output = 0 to 100°C Default Setting
- -50 to 120°C (-58 to 248°F) Full Programmable Range
M12TXSS
Depuis
C$
170.00

L'aperçu du produit
- Accuracy Class A
- Accuracy Note Class A per IEC60751
- Body Material Stainless steel AISI 316L
- Downscale 3.6 mA
- Electrical Output 4 to 20 mA
- EMC In accordance to EN 61326
- Enclosure Ratings IP67
- Enclosure Ratings Note IP65 and IP67 in accordance to IEC60529
- Factory Setting 0 to 100°C/ sensor break >21 mA (upscale)
- Operating Temperature, Max 80 °C
- Operating Temperature, Min -40 °C
- Permissible Load 700
- Permissible Load Note 700 @ 24 Vdc [RLo= (Vsupply-8.5)/0.020]
- Probe Diameter 3 mm
- Process Temperature, Max 120 °C
- Process Temperature, Min -50 °C
- Range Configurations Note It is possible set the input temperature range (span) by the M12TX-CONFIG-SETconfiguration kit (PC with OS Windows® required)
- Relative Humidity, Max 1
- Relative Humidity, Min 0%
- Response Time < 50 ms
- Response Time Note Diameter 3 mm (0.12") <3.5 seconds, diameter 6 mm (0.24") <13 seconds (test in water to IEC751- time for reaching 63.2% of instantaneous temperature change)
- RTD Type 100 Ω Pt Sensor
- Sensor Break Monitoring Note Selectable: Upscale (>21.0 mA) or downscale (<3.6 mA) action Sensor Short-Circuit: Fixed to downscale (<3.6 mA) action
- Sensor Error Compensation Note Over 2 points (maximum 1% of span)
- Span, Min 50°C
- Span Note 50°C (122°F) [if the zero value is set between one of these values: -40°C (-40°F), -20°C (-4°F), 0°C (32°F), 20°C (68°F), 40°C (104°F), the minimum span is 20°C (68°F) rather than 50°C (122°F)]
- Supply Power Effects Negligible
- Supply Voltage 8.5 to 32 Vdc
- Temperature Influence, Max ± 0.3%
- Temperature Influence Note Maximum of ± 0.3°C/25°C (77°F) or ± 0.3% of span/25°C (77°F)
- Temperature Sensor Type RTD
- Thermocouple Type N/A
- Tmax. Electronic 80°C
- Upscale 21 mA
- Zero Setting Adjustable Range Any value between -50 to 50°C (-58 to 122°F)
RTD 100 Ω sensor with built-in transmitter is programmable by a computer. The configuration kit required for programming is M12TX-CONFIG. This unique probe is ideal for areas with space limitations where traditional head connections are too large to fit. The M12 thread design offers a secure industrial connection.
Specifications
Body: Stainless steel AISI 316L
Probe Length: 13 mm (0.51') and 24 mm (0.94')
Diameter: à˜ 3 mm (0.1')
Probe Type: Thermowell AISI 316L SS
Probe Minimum Bending Radius: Three-times the outer diameter [except the sensing tip which length is 30 mm (1.2')]
Connection: 1/8 NPT male thread
Sensor: RTD PT100 class A up to 300°C in accordance to IEC751
Range: -50 to 120°C (-58 to 248°F)
Tmax. Electronic: 80°C (176°F)
Sensor Break Monitoring:
Selectable: Upscale (>21.0 mA) or downscale (<3.6 mA) action
Sensor Short-Circuit: Fixed to downscale (<3.6 mA) action
Output:
Signal: 4 to 20 mA
Permissible Load: 700 @ 24 Vdc [RLo= (Vsupply-8,5)/0,020]
Default Factory Setting for 4 to 20mA output: 0 to 100°C
Response Time (90%): <50 ms
Isolation In-Out: Non-isolated
Power Supply: 8.5 to 32 Vdc (polarity protected)
Environment Conditions:
Temperature: -40 to 80°C (-40 to 176°F) for plastic body
Relative Humidity: 0 to 100%
EMC: In accordance to EN 61326
Degree of Protection: IP65 and IP67 in accordance to IEC60529
Accuracy:
Transmitter: Maximum of ± 0.2°C or ± 0.2% of span
Sensor: Class A in accordance to IEC751
Temperature Influence [Deviation from 20°C (68°F)]: Maximum of ± 0.3°C/25°C (77°F) or ± 0.3% of span/25°C (77°F)
Supply Voltage Influence: Negligible
Range Configurations: It is possible set the input temperature range (span) by the M12TX-CONFIG configuration kit (PC with OS Windows required)
Zero Adjustments: Any value between -50 to 50°C
Minimum Span: 50°C (122°F) [if the zero value is set between one of these values: -40°C (-40°F), -20°C (-4°F), 0°C (32°F), 20°C (68°F), 40°C (104°F), the minimum span is 20°C (68°F) rather than 50°C (122°F)]
Sensor Error Compensation: Over 2 points (maximum 1% of span)
Factory Setting: 0 to 100°C/sensor break >21 mA (upscale)
Response Time: Diameter 3 mm <3.5 seconds, diameter 6 mm <13 seconds (test in water to IEC751- time for reaching 63.2% of instantaneous temperature change)
Specifications
Body: Stainless steel AISI 316L
Probe Length: 13 mm (0.51') and 24 mm (0.94')
Diameter: à˜ 3 mm (0.1')
Probe Type: Thermowell AISI 316L SS
Probe Minimum Bending Radius: Three-times the outer diameter [except the sensing tip which length is 30 mm (1.2')]
Connection: 1/8 NPT male thread
Sensor: RTD PT100 class A up to 300°C in accordance to IEC751
Range: -50 to 120°C (-58 to 248°F)
Tmax. Electronic: 80°C (176°F)
Sensor Break Monitoring:
Selectable: Upscale (>21.0 mA) or downscale (<3.6 mA) action
Sensor Short-Circuit: Fixed to downscale (<3.6 mA) action
Output:
Signal: 4 to 20 mA
Permissible Load: 700 @ 24 Vdc [RLo= (Vsupply-8,5)/0,020]
Default Factory Setting for 4 to 20mA output: 0 to 100°C
Response Time (90%): <50 ms
Isolation In-Out: Non-isolated
Power Supply: 8.5 to 32 Vdc (polarity protected)
Environment Conditions:
Temperature: -40 to 80°C (-40 to 176°F) for plastic body
Relative Humidity: 0 to 100%
EMC: In accordance to EN 61326
Degree of Protection: IP65 and IP67 in accordance to IEC60529
Accuracy:
Transmitter: Maximum of ± 0.2°C or ± 0.2% of span
Sensor: Class A in accordance to IEC751
Temperature Influence [Deviation from 20°C (68°F)]: Maximum of ± 0.3°C/25°C (77°F) or ± 0.3% of span/25°C (77°F)
Supply Voltage Influence: Negligible
Range Configurations: It is possible set the input temperature range (span) by the M12TX-CONFIG configuration kit (PC with OS Windows required)
Zero Adjustments: Any value between -50 to 50°C
Minimum Span: 50°C (122°F) [if the zero value is set between one of these values: -40°C (-40°F), -20°C (-4°F), 0°C (32°F), 20°C (68°F), 40°C (104°F), the minimum span is 20°C (68°F) rather than 50°C (122°F)]
Sensor Error Compensation: Over 2 points (maximum 1% of span)
Factory Setting: 0 to 100°C/sensor break >21 mA (upscale)
Response Time: Diameter 3 mm <3.5 seconds, diameter 6 mm <13 seconds (test in water to IEC751- time for reaching 63.2% of instantaneous temperature change)