PX276

Duct Pressure Transmitter with Field Selectable Ranges

  • Duct pressure transducer and static pressure probe integrated as one unit
  • Up to 3 field selectable ranges in a single unit
  • Rugged NEMA-4 enclosure with external mounting bracket
  • Short circuit and reverse polarity protected
Duct Pressure Transmitter with Field Selectable Ranges PX276 PX276
loading all models
L'aperçu du produit
  • Accuracy Note ± 1.0% FS
  • Additional Features 1 Rangable
  • Compensated Temperature, Max 65 °C
  • Compensated Temperature, Min -4 °C
  • Electrical Connection Screw Terminal
  • Enclosure Material Enamel coated 18GA steel
  • Enclosure Ratings IP65, NEMA Type 4
  • Media Compatibility Air, Argon, Carbon dioxide, Helium, Hydrogen, Nitrogen, Non-corrosive Dry Gasses, Oxygen
  • Pressure, Min 0 inH₂O
  • Thermal Effect 0.02% FS/°C
  • Thermal Effect on Span error, Max ± 1% gage
  • Weight 454 g
The PX276 is a rugged duct pressure transmitter that offers up to 3 field-selectable ranges in one unit. The PX276 incorporates a NEMA 4 (IP65) enclosure and a static pressure probe in one unit for ease of installation and accurate static duct pressure readings.

SPECIFICATIONS
Excitation: 12 to 40 Vdc
Output: 4 to 20 mA (2 wire)
Accuracy: ± 1.0% FS
Operating Temperature: -18 to 80°C (0 to 175°F)
Compensated Temperature: -4 to 65°C (25 to 150°F)
Thermal Effects: 0.02%URL/°C (± 0.0125% URL/°F)
Proof Pressure: 10 psi
Media Compatibility: Clean, dry air or inert gas
Termination: Screw terminal block
Enclosure: Enamel-coated, 18 GA steel enclosure with 6061T aluminum probe; NEMA 4 (IP65) rating
Weight: 454 g (1.0 lb)
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 4ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 29ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
fr-CA
fr
Enregistrer la configuration