Bienvenue sur notre nouveau site Web, vous avez rencontré un problème ou un bug? Veuillez le signaler ici

CL3515R-Series
Simulateur/thermomètre thermocouple portatif avec USB{USB}{RS232}{WINXP}
- USB et interface RS232 avec logiciel Windows
- 11 types de thermocouples: K, J, T, E, R, S, N, L, U, B, et C (en °C, °F, ou K), et sortie signal mV
- 2 Chaînes pour mesures de température thermocouple (T1 et T2) avec ajustement Offset
- Sélecteur rotatif ergonomique permet l'opération avec une main
CL3515R-Series
Depuis
C$
607.28

L'aperçu du produit
- Accuracy ± 0.05 %
- Accuracy Note ± (0.05% of reading + 0.5 °C)
- Additional Features 1 NIST Calibration Certificate
- Battery life 17 hr
- Display Type LCD
- Display Type Note 4-digit LCD with maxreading of 19999
- Electrical Connection Miniature Female Thermocouple Connector
- Height 192 mm
- Length 52.5 mm
- Milli Volt Accuracy ± 0.025% + 1 digit
- Milli Volt Resolution 10 µV
- Millivolt Range -25 to 75 mV
- Number of Channels 2
- Operating Temperature, Min 0 °C
- Relative Humidity, Max 0.7
- Relative Humidity, Min 0%
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -20 °C
- Storage Temperature Note -20 to 60°C (-4 to 140°F ),<70%RH
- Supply Voltage 9 Vdc
- Temperature Coefficient 0.1 times the applicable accuracy specification per °C from 0 to 18°C and 28 to 50°C (32 to 64°F and 82 to 122°F)
- Temperature Measurement, Max 2316 °C
- Temperature Measurement, Min 0 °C
- Thermocouple Simulation Resolution Range Thermocouple Simulation Range: Resolution: 0.1° (1° for R/S/B/C); Accuracy*: ± (0.3°C + 10 µV)
- Thermocouple Type J, K, T, E, R, S, B, C, N, L, U
- Type of RTD Signal None
- Weight 320 g
- Width 91 mm
Description
Le CL3515R est un calibrateur/thermomètre avec un affichage rétroéclairé à 4 digits. Il a été conçu pour mesurer et simuler une large plage de types de thermocouples: K/J/T/E/R/S/N/L/U/B/C. Il y a aussi une plage de tension de sortie de -25mV à + 75mV.Le thermomètre présente une double entrée thermocouple et un offset thermocouple ajustable.
CARACTERISTIQUES
Thermocouple (Résolution) plage de mesure*:
K: (0,1°) -200 à 1372°C
J: (0,1°) -210 à 1200°C
T: (0,1°) -250 à 400°C
E: (0,1°) -250 à 1000°C
R: (1°) 0 à 1767°C
S: (1°) 0 à 1767°C
N: (0,1°) -200 à 1300°C
L: (0,1°) -200 à 900°C
U: (0,1°) -200 à 600°C
B: (1°) 600 à 1820°C
C: (1°) 0 à 2316°C
* Selon la température standard ITS-90.
Précision:
Thermocouple types K/J/T/E/L/U:
±(0,05% lec + 0,5°C) -50 à 1372°C
±(0,05% lec + 1,0°C) -50 à -250°C
Thermocouple type N:
±(0,05% lec + 1,0°C) -50 à 0°C
±(0,05% lec + 0,5°C) 0 à 1300°C
Thermocouple types R/S/B/C:
±(0,05% lec + 2°C) 0 à 1767°C
Plage de simulation de thermocouple:
Résolution: 0,1° (1° pour R/S/B/C)
Précision*: ±(0,3°C + 10 µV)
* Précision est spécifiée pour les températures de fonctionnement de 18 à 28°C pedant 1 an, ne comprend pas des erreurs du thermocouple.
Plage mV : -25 à 75 mV
Résolution: 10 µV
Précision: ±0,025% + 1 digit
Coefficient de température: 0,1 fois la spécification de précision par °C de 0 à 18°C et 28 à 50°C
Affichage: 4 digits, rétroéclairé avec lecture maxi de 19999
Connecteur entrée: prise thermocouple miniature (SMP) Alimentation: batterie 9V (comprise)
Autonomie batterie: 17,5 heures, nominale avec indicateur batterie faible
Dimensions: H 192 x L 91 x D 52,5 mm
Poids: 320 g
Plages de fonctionnement ambiantes environnementales: 0 à 50°C, <80% HR
Plages de stockage: -20 à 60°C, <70% HR
PDF et manuels
Accessories
Afficher les notes et les avis
Rated 2 out of
5
by
Arnold from
OK Quality 2 channel calibrated unit
We bought this because it is calibrated. Seems susceptible to EMI. The HVAC system combined with other adjacent CE compliant test equipment cause this to fail the self test. Other thermometers and test equipment do not fail under the same conditions. It is good that it reports the errors, but it would be nice if it was more robust. Good tech support to identify EMI as the cause.
Date published: 2017-06-27
Rated 1 out of
5
by
Facilities guy from
Does not work with Windows 10
This product will not work with Windows 10 and no relief in sight...
Date published: 2018-12-06
Rated 4 out of
5
by
bhelmet from
Meets my needs
I still have a lot to learn about all the capabilities of this equipment, but it works for my calibration needs. I would like to see more documentation on the capabilities and how to use them. What I have seen thus far is a bit lacking for me. I haven't had the opportunity to play with it outside of performing my calibrations. However, it is great for my calibration needs
Date published: 2018-10-25
Rated 5 out of
5
by
Erich from
Impressive Accuracy
I have purchased 3 of these over the past couple of years for training customers on properly performing SAT's and they work great. They are every bi as accurate as $2,000 calibrators we have used in the past and they handle a wide range of thermocouples.
Date published: 2017-10-10
Rated 1 out of
5
by
Harvey from
Rarely in specification
Always having to adjust and issuing out of tolerances. Get what you pay for, buy a Fluke.
Date published: 2019-10-08
Rated 5 out of
5
by
Iain C from
1
works as intended to calibrate plc cards. Easy to use.
Date published: 2025-01-24
Rated 3 out of
5
by
JSpiker from
Works, but not great
It does produce a TC signal that is accurate and adjustable. There are 2 outputs, but they do not seem to be independent, they both put out the same signal. The temperature logging is kind of confusing and it is easy to get stuck in an unknown state. It also beeps very loudly each time you press a button.
Date published: 2024-03-21
Rated 5 out of
5
by
MWong from
Staff Mechanical Engineer
Simple and straightforward usage with multiple thermocouple type input accepted.
Date published: 2023-02-27
Can I communicate with the unit through USB in my own program instead of the windows program. I am trying to use it under Mac so would like to open it as a serial port to control it. Is there any documents of Remote Programming Commands for the CL3515R?
Hi Apple
We do have command protocol for the CL3515R. Please email Temp@omega.com to request copy.
Best regards
OMEGA
Date published: 2020-04-03
What is the expanded uncertainty considering k=2 in degree C for the ranges of 0-100, 100-200, 200-300 and 300-400 degrees, thermoucouple type J. I could not find the information over the manual.
Hi MB0723,
Thank you for the inquiry. This product is provided with stated accuracies only. Uncertainties are not provided unfortunately.
Best regards
OMEGA
Date published: 2023-07-31
Since most modern computers don't have disc drives, where can I download the software? I don't see it on the website.
Scroll to the bottom of the page. Under "Quick Links", there is a section called "Software". That will take you to Omega's web page containing all of their software. For this specific item, click on "Temperature", then find this model number on the FTP site.
Date published: 2024-05-07
Can this unit be used to calibrate heat controllers? I cannot see the receptacles or this information within the description.
Thank you for the inquiry. As long as the meter or controller is using the same thermocouple type as the controller, then the CL3515R can be used to verify the accuracy of the thermocouple reading.
Best regards
OMEGA
Date published: 2024-11-23
I am having an issue getting the software to see the calibrator. When I plug it in and try to select a port it does not pull up an option to click anything. Any suggestions? Also is there any more information on how to calibrate than what's in the manual?
Please make sure you are selecting Run As Administrator to install the USB driver. Check the Device Manager to confirm the USB driver is correctly loaded and Windows assign a port number.
Date published: 2023-02-08
Does this unit come with the optical to USB interface cable, if not what is the P/N for the cable?
Thanks for connecting with Omega. The product manager has confirmed that this product does come with a USB cable.
Date published: 2023-02-08
What is the refresh rate?
Hi Gior,
Thank you for the inquiry. Refresh rate is once per second.
Best regards
OMEGA
Date published: 2022-04-06
Does Omega offer a calibration/re-certification service for the CL3515R, and what is the suggested frequency of recalibration?
The CL3515R can be re-calibrated and it is recommended it be done once per year. This can be set up by the Customer Service group when you are ready to move forward.
Date published: 2022-04-18