Thermomètre à Thermocouple de Haute Précision, Interface sans fils ou USB {USB}{wireless}{WINXP}HH802-803HH802-803
L'aperçu du produit
Additional Features 1
MIN/MAX/REL/AVG/HOLD, Auto power off, Hi-Lo Limit alarm
Battery Size and Type
Alkaline AAA
Display Type
LCD
Display Type Note
Dual 5 digit + Elapsed Time
Form factor
Handheld
Height
38 mm
Length
160 mm
Measurement Type
Temperature
Number of Measurements
2
Operating Temperature, Max
50 °C
Operating Temperature, Min
0 °C
Sampling Rate
1 Hz (1 update/second)
Storage Humidity, Max
0.8
Storage Humidity, Min
0
Storage Temperature, Max
60 °C
Storage Temperature, Min
-20 °C
Storage Temperature Note
-20 to 60°C, 0 to 80% RH with battery removed
Supply Power Type
Battery
Temperature Accuracy
± 0.05 %
Temperature Accuracy Note
Stated accuracy at 23°C
± 5°C, <75% RH
± (0.05% rdg + 0.3°C) -50 to 1370°C
± (0.05% rdg + 0.7°C) -50 to -200°C
± (0.05% rdg + 0.6°F) -58 to 2498°F
± (0.05% rdg + 1.4°F) -58 to -328°F
Temperature Measurement, Max
1370 °C
Temperature Measurement, Min
-200 °C
Temperature Sensor Type
Thermocouple
Weight
270 g
Weight Note
Approx. 270 g (0.5 lb) including batteries
Width
83 mm
Description
Le thermomètre à thermocouple HH802U est un thermomètre de haute précision, qui accepte un thermocouple type J ou K. Il dispose d'un double affichage à 5 digits pour mesurer la température et un troisième affichage du temps écoulé. Ce modèle comprend un port USB qui permet une connexion directe entre l'appareil et un PC (câble USB et logiciel sous Windows vendus séparément). Il comprend également une prise d'alimentation CC pour une utilisation prolongée avec l'adaptateur AC (vendu séparément). Le modèle HH802WE dispose une interface sans fil qui remplace l'USB et permet le transfert des données vers un PC, sans câble (le logiciel sous Windows et récepteur sans fil à acheter séparément). Le modèle HH802U dispose des fonctions de rappel des lectures de REL, de rappel de limite de Hi/Lo et de compensation de soudure froide.
What is included to export data, software, usb cable?
Asked by: CarlR
To export data you will need to purchase the HH800-SW which will include the Software and USB communication cable.
Answered by: ANBENG
Date published: 2020-04-03
Does this model store data on it, or is the separate Windows program required in order to log the data over time?
Asked by: DSmart
No internal memory for storage is available for this model. Requires software running on a PC to log data.
Answered by: ANBENG
Date published: 2020-04-03
Are the thermocouple probes included or is it just the data logger?
Asked by: Anusha
Thank you for your inquiry. The HH802U comes with a ready made beaded Type K thermocouple (one per channel).
Answered by: NorwalkAE_SP
Date published: 2024-04-25
Does this device comes with calibration certificate?
Asked by: HamidM
The HH802U comes with a NIST Certificate without temperature points.
Answered by: Applications Engineer ANB
Date published: 2023-02-13
On the Omega HH802W Thermometer, how do I:
1. Turn Off the Auto Power Off (APO) mode?
2. Turn On the APO mode?
3. Adjust the time of the APO mode?
Asked by: Jack 713
To turn off the APO you need to hold the power button for > 6 seconds and it will indicate on display APO off. To turn it back on shut the meter off and repower after a few seconds and there is no time adjustment for APO.
Answered by: ApplicationEngineerANB
Date published: 2024-06-03
What type of thermocouple connector (type K) does this unit accept? Is it a standard size or a miniature thermocouple connector?
Asked by: Klovus
The HH802W accepts a male miniature Thermocouple connector with the flat blades.
Answered by: ApplicationEngineerANB
Date published: 2023-12-04
Optional AC Adaptor
Where can I found this?
Asked by: smurf
Thank you for your inquiry. Please click link https://www.omega.com/en-us/sensors-and-sensing-equipment/temperature/hygrothermometers/p/HH800-ADAPTOR
Answered by: Application Engineer CB
Date published: 2020-11-11
What could be the problem of " Time Out" occur during the temperature recording?
Asked by: Amornrat
A Time Out error usually indicates communication connection was lost between the meter and PC during recording. Check for loose fittings, bad communication cable or low battery on the meter as possible culprits.