RTD Temperature Transmitter - 316 Stainless Steel Construction
- Ideal for Wash-down, Outdoor, Sanitary or Marine Environments
- Compact Design Attaches Directly to Most 3-Wire and 4-Wire, Pt100, 0.00385 Curve Sanitary Probes and Sensors with M12 Style Connectors
- True 2-Wire Operation, Loop Powered 4 to 20 mA Output
- Sealed Signal Conditioner Built Into NEMA 4X (IP67) 316 Stainless Steel Housing

Product Overview
- Accuracy ± 0.1 %
- Approvals CE
- Bandwidth 3 Hz (-3 dB)
- Diameter 38 mm
- Electrical Connection M12 4 pin Male
- Electrical Output 4 to 20 mA
- Enclosure Ratings IP67, NEMA 4
- Hysteresis ± 0.1%
- Input Electrical Connection M12 4 pin Female
- Linearity ± 0.1%
- Loop Resistance, Max (Vs - 8V) / 0.020 A
- Materials 316 Stainless Steel
- Mounting Mounting Bracket
- Number of Wires 2
- Operating Temperature, Max 85 °C
- Operating Temperature, Min -40 °C
- Output Signal Current
- Repeatability ± 0.1%
- Span Adjustment ± 25 %
- Supply Power 8 to 35 Vdc
- Supply Power Type DC
- Temperature Coefficient α = 0.00385 curve
- Zero Setting Adjustable Range ± 25%
Loop Powered! The OMEGA™ 2-wire TX94A-SS Series temperature transmitters are high-performance, industrial transmitters designed for direct connection to most CIP (clean-inplace) Sanitary Pt100 probes and sensors that incorporate a M12 style connection. All models feature a miniature signal conditioner built into a sealed stainless steel connector housing. The signal conditioner converts the resistive change of a 100 Ω, 0.00385 RTD probe or sensor into an industry-standard 2-wire, 4 to 20 mA analog output across a dedicated temperature range. Output connection is also a standard M12 connector.
M12 RTD SENSOR WIRING | |
![]() | ![]() |
Wiring option #1 | Wiring option #2 |
The TX94A-SS transmitter is provided with two RTD wiring configurations as shown. The wiring style is identified at the end of the model number by either a "-1" or "-2". If used with an Omega PR-21 sensor, then the "-1" or "-2" selection should match the final charecters following the "-M12" in the sensor model number (ie: PR-21A-3-100-A-1/4-0600-M12-1). When using the transmitter with an Omega PR-22 style RTD, then please select the -2 wiring configuration.
Examples:
For use with a PR-21A-3-100-A-1/4-0600-M12-1, use an TX94A-2-SS-M12F-1 Transmitter (Range is -18 to 93°C).
For use with a PR-22-3-100-A-M6-250-M12 sensor, use an TX94A-2-SS-M12F-2 Transmitter (Range is -18 to 93°C).
SPECIFICATIONS
Supply Voltage: 8 to 35 Vdc; (24 Vdc recommended)
Max Load: Rmax (Ω) = (Vsupply - 8V)/0.02 A
Output: Linearized 4 to 20 mA
Accuracy: ± 0.1% FS (includes effects of linearity, hysteresis and repeatability)
Frequency Response: 3 dB @ 3Hz
Range:
1: -40 to 49°C (-40 to 120°F)
2: -18 to 93°C (0 to 200°F)
3: -18 to 149°C (0 to 300°F)
4: -18 to 260°C (0 to 500°F)
Input: 3-wire, Pt100 (α = 0.00385), 4-wire RTD compatible Probe/Sensor Input
Connection:
TX94A-SS-M12M: M12 style, 4-pin male
TX94A-SS-M12F: M12 style, 4-pin female
Transmitter Operating Temp: -40 to 85°C (-40 to 185°F)
Output Connection (All Models): M12 style (order cable from table below)
Housing: 316 stainless steel, polished mirror finish
Approvals: CE marked
Dimensions: 51 H x 38 mm D (2.0 x 1.50")